Author:
Funderburgh L J,Zipf W B,Sotos J F
Abstract
Abstract
We evaluated a commercial (Radioassay Systems) radioimmunoassay kit for testosterone in which a double-antibody method and an 125I tracer are used. The 50-microL serum sample is assayed directly with no pre-extraction step. The assay system has good sensitivity (0.01 microgram/L), good precision (CV 5.2%), and little cross reactivity between testosterone and other androgens. Results compared well with those by a tritiated radioimmunoassay involving extraction (r = 0.98). Values obtained for 97 children compared well with published pediatric normal values. Evidently, this kit provides a rapid and reliable measurement of testosterone in serum specimens and is acceptable for use with pediatric patients.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献