Affiliation:
1. The University of Reading, Department of Agriculture, Faculty Analytical Laboratory, PO Box 236, Reading RG6 6AT, UK
Abstract
Abstract
The new Soxflo instrument was evaluated for the determination of crude fat in foods and animal feeds. Samples are packed into small columns and extracted with petroleum ether at room temperature. The Soxflo yielded accurate data from foods, ranging from 0.4 to 73.2% crude fat, compared with Soxhlet extractions and Certified Reference Materials, for which recoveries averaged 99.7 and 100.7%, respectively. Relative standard deviations (1.81 %) were approximately half those of Soxhlet extractions (3.68%). Regression analysis of the data suggested that there was no proportional bias. A small but acceptable constant bias was measured. Soxflo extractions are easy to perform and take approximately 1 h to complete. The main difference between the Soxflo and Soxhlet techniques lies in the extraction procedure. Estimated savings during extractions are in time (85% reduction), energy (95%), cooling water (100%), and solvents (50%). Soxflo extractions are, therefore, more environmentally friendly than Soxhlet extractions.
Publisher
Oxford University Press (OUP)
Subject
Pharmacology,Agronomy and Crop Science,Environmental Chemistry,Food Science,Analytical Chemistry
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献