Design of a new breast vacuum bag to reduce the global and local setup errors and to reduce PTV margin in post-mastectomy radiation therapy

Author:

Shen Konglong1,Xiong Jie2,Wang Zhiguo1,Wang Weifeng3,Li Wan1,Zhou Jidan1,Deng Zhonghua1,Li Bin1,Zhong Renming1

Affiliation:

1. Department of Radiotherapy, Division of Radiation Physics, State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, P.R. China

2. Radiology Department, West China Hospital, Sichuan University, Chengdu 610041, P.R. China

3. Jingzhou Central Hospital, the Second Clinical Medical College, Yangtze University, Jinzhou 121001, P.R. China

Abstract

Abstract To design a new breast vacuum bag to reduce global and local setup errors in post-mastectomy radiation therapy (PMRT). A total of 24 PMRT patients were immobilized with an old vacuum bag and 26 PMRT patients were immobilized with a new vacuum bag. The registration results were analysed using four regions of interest (ROI): the global ROI [including the whole region of the planning target volume (PTV), GROI], the supraclavicular area (SROI), the ipsilateral chest wall region (CROI) and the ipsilateral arm region (AROI). The global and local setup errors of the two groups were compared. The global setup errors of the new vacuum group were significantly smaller than those in the old vacuum group with the exception of yaw axes (P < 0.05). The systematic error (Σ) and random error (σ) ranged from 1.21 to 2.13 mm. In the new vacuum group, the local setup errors in the medial-lateral (ML) direction and roll axes for CROI (the Σ and σ ranged from 0.65 to 1.35 mm), and the local setup errors in ML and superior-inferior (SI) directions for SROI were significantly smaller than those in the old vacuum group. The total required PTV margins for the chest wall in ML, SI, and anterior-posterior (AP) were 4.40, 3.12 and 3.77 mm respectively. The new vacuum bag can significantly reduce the global setup errors and local setup errors in PMRT. The respiratory motion of the chest wall was negligible, and the 5 mm PTV margin could cover the local setup errors in PMRT using the new vacuum bag with cone beam CT (CBCT) correction.

Funder

Science and Technology Support Program of Sichuan Province

Publisher

Oxford University Press (OUP)

Subject

Health, Toxicology and Mutagenesis,Radiology Nuclear Medicine and imaging,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3