Intraoperative localization of cardiac conduction tissue regions using real-time fibre-optic confocal microscopy: first in human trial

Author:

Kaza Aditya K1,Mondal Abhijit1ORCID,Piekarski Breanna1,Sachse Frank B23ORCID,Hitchcock Robert2ORCID

Affiliation:

1. Department of Cardiac Surgery, Boston Children’s Hospital, Harvard Medical School, Boston, MA, USA

2. Department of Biomedical Engineering, University of Utah, Salt Lake City, UT, USA

3. Nora Eccles Harrison Cardiovascular Research and Training Institute, University of Utah, Salt Lake City, UT, USA

Abstract

Abstract OBJECTIVES The aim of this study was to evaluate the feasibility and safety of fibre-optic confocal microscopy (FCM) using fluorescein sodium dye for the intraoperative location of conduction tissue regions during paediatric heart surgery. METHODS The pilot study included 6 patients undergoing elective surgery for the closure of isolated secundum atrial septal defect aged 30 days to 21 years. FCM imaging was integrated within the normal intraoperative protocol for atrial septal defect repair. Fluorescein sodium dye was applied on the arrested heart. FCM images were acquired at the atrioventricular node region, sinus node region and right ventricle (RV). Total imaging time was limited to 3 min. Any adverse events related to the study were recorded and analysed. Subjects received standard postoperative care. Trained reviewers (n = 9) classified, de-identified and randomized FCM images (n = 60) recorded from the patients as presenting striated, reticulated or indistinguishable microstructures. The reliability of reviewer agreement was assessed using Fleiss’ kappa. RESULTS The FCM imaging instruments were integrated effectively into the cardiac surgery operating room. All adverse events found in the study were deemed expected and not related to FCM imaging. Reticulated myocardial microstructures were found during FCM imaging at atrioventricular node and sinus node regions, while striated microstructures were observed in RV. Reliability of agreement of reviewers classifying the FCM images was high (Fleiss’ kappa: 0.822). CONCLUSIONS FCM using fluorescein sodium dye was found to be safe for use during paediatric heart surgery. The study demonstrates the potential for FCM to be effective in identifying conduction tissue regions during congenital heart surgery. Clinical trial registration number NCT03189134.

Funder

National Institutes of Health

NIH

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,General Medicine,Surgery

Cited by 10 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Innovation and areas for future research;Nadas' Pediatric Cardiology;2025

2. Emerging techniques in cardiac surgery;Nadas' Pediatric Cardiology;2025

3. Intraoperative characterization of cardiac tissue: the potential of light scattering spectroscopy;Journal of Biomedical Optics;2024-06-05

4. Advances in the Congenital Heart Disease;Japanese Journal of Cardiovascular Surgery;2023-11-15

5. Conduction in congenital heart surgery;The Journal of Thoracic and Cardiovascular Surgery;2023-10