A case report of inappropriate inhibition of ventricular pacing due to a unique pacemaker electrogram storage feature

Author:

Khan Habib R1ORCID,Chan William K1ORCID,Kanawati Juliana1ORCID,Yee Raymond1ORCID

Affiliation:

1. London Health Sciences Centre, University of Western Ontario, 339 Windermere Road, London, ON N6A 5A5, Canada

Abstract

Abstract Background Modern permanent pacemakers (PPMs) have individual features designed to identify cardiac rhythm abnormalities and improve their performance. Inappropriate pacing inhibition may be an undesired outcome from these features and cause symptoms in patients who require frequent pacing, leading to dizziness, and syncope. Inappropriate inhibition can be difficult to identify in circumstances that are intermittent and difficult to reproduce. Case summary A 57-year-old female underwent a mitral valve replacement (MVR) for severe mitral stenosis. One month following MVR, she presented with symptomatic third-degree atrioventricular block, and a dual-chamber PPM (Advisa™, Medtronic, Minneapolis, USA) was implanted and programmed DDD 50–130 b.p.m. At the 3-month follow-up, she reported frequent episodes of lightheadedness. She was found to have intermittent ventricular pacing inhibition on a 48-h Holter monitor due to an internal function of the Advisa™ series of PPMs that attempts to store an electrogram (EGM) every 1 h and 30 s. During the EGM storage, an amplified signal from the storage capacitor can result in oversensing by the ventricular channel and inappropriate pacing inhibition. Discussion To rectify the issue, the ventricular lead sensitivity value was increased from 0.9 mV to 1.2 mV. No instances of inappropriate ventricular pacing inhibition were noted on follow-up. To our knowledge, this is a rare case of inappropriate ventricular pacing inhibition caused by a combination of PPM self-adjusting sensitivity algorithm and oversensing every 1 h and 30 s from an amplified storage capacitor. Physicians should be aware of this possible complication and differentiate it from device or lead malfunction.

Publisher

Oxford University Press (OUP)

Subject

Cardiology and Cardiovascular Medicine

Reference10 articles.

1. Pacing features that mimic malfunction: A review of current programmable and automated device functions that cause confusion in the clinical setting;Lloyd;J Cardiovasc Electrophysiol,2009

2. A review of specialized and automated features in implantable cardiac devices;Ayinde;J Cardiovasc Electrophysiol,2020

3. 2018 ACC/AHA/HRS Guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines, and the Heart Rhythm Society;Kusumoto;Circulation,2019

4. 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA);Brignole;Eur Heart J,2013

5. Double chamber cross-talk, the ultimate double-cross;Almehmadi;Pacing Clin Electrophysiol,2020

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3