Abstract
Abstract
In screening mammography, the automatic exposure control (AEC) device selects the optimum exposure parameters for each patient exposure. In practice, this device is calibrated to deliver an optimum figure of merit (FOM) for each set of exposure parameters. This study utilises the central composite design methodology to verify the operating level of the AEC performance for polymethyl methacrylate (PMMA) thicknesses of 4 and 5 cm. Quadratic surface plots reveal that the exposure parameters of the default AEC mode are optimised for the 4-cm thickness of the PMMA. However, the exposure parameters of the 5-cm thickness of PMMA need fine-tuning to obtain a maximum value for the FOM.
Funder
South African Medical Research Council
Economic Competitiveness and Support Package
Medical Research Council