Affiliation:
1. Department of Defense Biotechnology High Performance Computing Software Applications Institute, Telemedicine and Advanced Technology Research Center, United States Army Medical Research and Development Command , Fort Detrick, MD , USA
2. The Henry M. Jackson Foundation for the Advancement of Military Medicine, Inc. , Bethesda, MD , USA
Abstract
Abstract
Study Objectives
Wearable sleep-tracker devices are ubiquitously used to measure sleep; however, the estimated sleep parameters often differ from the gold-standard polysomnography (PSG). It is unclear to what extent we can tolerate these errors within the context of a particular clinical or operational application. Here, we sought to develop a method to quantitatively determine whether a sleep tracker yields acceptable sleep-parameter estimates for assessing alertness impairment.
Methods
Using literature data, we characterized sleep-measurement errors of 18 unique sleep-tracker devices with respect to PSG. Then, using predictions based on the unified model of performance, we compared the temporal variation of alertness in terms of the psychomotor vigilance test mean response time for simulations with and without added PSG-device sleep-measurement errors, for nominal schedules of 5, 8, or 9 hours of sleep/night or an irregular sleep schedule each night for 30 consecutive days. Finally, we deemed a device error acceptable when the predicted differences were smaller than the within-subject variability of 30 milliseconds. We also established the capability to estimate the extent to which a specific sleep-tracker device meets this acceptance criterion.
Results
On average, the 18 sleep-tracker devices overestimated sleep duration by 19 (standard deviation = 44) minutes. Using these errors for 30 consecutive days, we found that, regardless of sleep schedule, in nearly 80% of the time the resulting predicted alertness differences were smaller than 30 milliseconds.
Conclusions
We provide a method to quantitatively determine whether a sleep-tracker device produces sleep measurements that are operationally acceptable for fatigue management.
Funder
U.S. Army Medical Research and Development Command
Publisher
Oxford University Press (OUP)
Subject
Physiology (medical),Neurology (clinical)
Reference30 articles.
1. Wearable sleep technology in clinical and research settings;de Zambotti;Med Sci Sports Exerc.,2019
2. Prospective validation of 2B-Cool: Integrating wearables and individualized predictive analytics to reduce heat injuries;Laxminarayan;Med Sci Sports Exerc.,2023
3. Direct comparison of two new actigraphs and polysomnography in children and adolescents;Meltzer;Sleep.,2012
4. Validation of Zulu watch against polysomnography and actigraphy for on-wrist sleep-wake determination and sleep-depth estimation;Devine,2020
5. A validation study of a commercial wearable device to automatically detect and estimate sleep;Miller;Biosensors,2021
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献