Creep adjustment of strain gauges based on granular NiCr-carbon thin films
-
Published:2021-03-12
Issue:1
Volume:10
Page:53-61
-
ISSN:2194-878X
-
Container-title:Journal of Sensors and Sensor Systems
-
language:en
-
Short-container-title:J. Sens. Sens. Syst.
Author:
Mathis Maximilian,Vollberg Dennis,Langosch Matthäus,Göttel Dirk,Lellig Angela,Schultes Günter
Abstract
Abstract. An important property of high-precision mechanical sensors such as force transducers or torque sensors is the so-called creep error. It is defined as the signal deviation over time at a constant load. Since this signal deviation results in a reduced accuracy of the sensor, it is beneficial to minimize the creep error. Many of these sensors consist of a metallic spring element and strain gauges. In order to realize a sensor with a creep error of almost zero, it is necessary to compensate for the creep behavior of the metallic spring element. This can be achieved by creep adjustment of the used strain gauges. Unlike standard metal foil strain gauges with a gauge factor of 2, a type of strain gauges based on sputter-deposited NiCr-carbon thin films on polymer substrates offers the advantage of an improved gauge factor of about 10. However, for this type of strain gauge, creep adjustment by customary methods is not possible. In order to remedy this disadvantage, a thorough creep analysis is carried out. Five major influences on the creep error of force transducers equipped with NiCr-carbon thin-film strain gauges are examined, namely, the material creep of the metallic spring element (1), the creep (relaxation) of the polymer substrate (2), the composition of the thin film (3), the strain transfer to the thin film (4), and the kind of strain field on the surface of the transducer (5). Consequently, we present two applicable methods for creep adjustment of NiCr-carbon thin- film strain gauges. The first method addresses the intrinsic creep behavior of the thin film by a modification of the film composition. With increasing Cr content (at the expense of Ni, the intrinsic negative creep error can be shifted towards zero. The second method is not based on the thin film itself but rather on a modification of the strain transfer from the polyimide carrier to the thin film. This is achieved by controlled cutting of well-defined deep trenches into the polymer substrate via a picosecond laser.
Publisher
Copernicus GmbH
Subject
Electrical and Electronic Engineering,Instrumentation
Reference19 articles.
1. Adhi, K. P., Owings, R. L., Railkar, T. A., Brown, W., and Malshe, A.: Chemical
modifications in femtosecond ultraviolet (248 nm) excimer laser
radiation-processed polyimide, Appl. Surf. Sci., 225, 324–331, 2004. a 2. Burfeindt, H., Gerlach, H.-J., and Will, R.: Foil strain gauge for a measuring
transducer, Patent No. EP 1 033 561 B1, Hottinger Messtechnik Baldwin, Germany, 2000. a 3. Cerino, M., Göttel, D., Probst, A.-C., and Schultes, G.: Schichtwiderstand mit
einem kohlenstoffhaltigen Widerstandsmaterial und Verfahren zu dessen
Herstellung, Patent No. DE 10 2015 006 057 A1, Hochschule für Technik und Wirtschaft des Saarlandes, Saarbrücken, Germany, ZeMa Zentrum für Mechatronik und Atomatisierungstechnik gemeinnützige GmbH, Saarbrücken, Gemrany, 2015. a, b 4. Gomez, D., Goenaga, I., Lizuain, I., and Ozaita, M.: Femtosecond laser ablation
for microfluidics, Opt. Eng., 44, 051105, https://doi.org/10.1117/1.1902783, 2005. a 5. Grange, H., Maeder, C., Bieth, C., Renard, S., and Delapierre, G.: Thin film
strain gauges on polymers: main characteristics, Sensor Actuat. A-Phys., 46, 213–217, 1995. a
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
|
|