An automated gas chromatography time-of-flight mass spectrometry instrument for the quantitative analysis of halocarbons in air

Author:

Obersteiner F.ORCID,Bönisch H.ORCID,Engel A.ORCID

Abstract

Abstract. We present the characterization and application of a new gas chromatography-time-of-flight mass spectrometry instrument (GC-TOFMS) for the quantitative analysis of halocarbons in air samples. The setup comprises three fundamental enhancements compared to our earlier work (Hoker et al., 2015): (1) full automation, (2) a mass resolving power R = m/Δ m of the TOFMS (Tofwerk AG, Switzerland) increased up to 4000 Th/Th and (3) a fully accessible data format of the mass spectrometric data. Automation in combination with the accessible data allowed an in-depth characterization of the instrument. Mass accuracy was found around 5 ppm after automatic recalibration of the mass axis in each measurement. A TOFMS configuration giving R = 3500 was chosen to provide an R-to-sensitivity ratio suitable for our purpose. Calculated detection limits were as low as a few femtograms as mass traces could be made highly specific for selected molecule fragments with the accurate mass information. The precision for substance quantification was 0.15 % at the best for an individual measurement and in general mainly determined by the signal-to-noise ratio of the chromatographic peak. The TOFMS was found to be linear within a concentration range from about 1 pg to 1 ng of analyte per Liter of air. At higher concentrations, non-linearities of a few percent were observed (precision level: 0.2 %) but could be attributed to a potential source within the detection system. A straight-forward correction for those non-linearities was applied in data processing, again by exploiting the accurate mass information. Based on the overall characterization results, the GC-TOFMS instrument was found to be very well-suited for the task of quantitative halocarbon trace gas observation and a big step forward compared to scanning, low resolution quadrupole MS and a TOFMS technique reported to be non-linear and restricted by a small dynamical range.

Publisher

Copernicus GmbH

Reference38 articles.

1. Arnold, T., Muhle, J., Salameh, P. K., Harth, C. M., Ivy, D. J., and Weiss, R. F.: Automated measurement of nitrogen trifluoride in ambient air, Anal. Chem., 84, 4798–4804, 2012.

2. Bassford, M. R., Simmonds, P. G., and Nickless, G.: An automated system for near-real-time monitoring of trace atmospheric halocarbons, Anal. Chem., 70, 958–965, 1998.

3. Bonk, T. and Humeny, A.: MALDI-TOF-MS analysis of protein and DNA, Neuroscientist, 7, 6–12, 2001.

4. Carpenter, L. J., Reimann, S., Burkholder, J. B., Clerbaux, C., Hall, B. D., Hossaini, R., Laube, J. C., and Yvon-Lewis, S. A.: Ozone-Depleting Substances (ODSs) and other gases of interest to the Montreal Protocol, in: Scientific Assessment of Ozone Depletion: 2014, Global Ozone Research and Monitoring Project – Report No. 55, 416 pp., World Meteorological Organization, Geneva, Switzerland, 2014.

5. Clemons, C. A. and Altshuller, A. P.: Responses of electron-capture detector to halogenated substances, Anal. Chem., 38, 133–136, 1966.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3