Abstract
In this study, piezoelectric micropumps with partial throw-away design have been developed and discussed. The actuator, a rectangular piezoelectric plate, can be separated from the pump body and be used repeatedly while the pump chamber is single-use to prevent infection. Thus, a low cost and waste solution for biomedical uses is achieved. A mechanical model is used to analyze the behavior of pump diaphragm and chamber geometry. Pump chamber with 20mm diameter is used due to its moderate elasticity. Then, 5 pumps with different depth of pump chamber have been made. The maximum flow rate, 5.7g/min, is reached by using a 0.5mm pump chamber with a 200V actuated voltage.
Publisher
Trans Tech Publications, Ltd.
Reference11 articles.
1. S. Böhm,W. Olthuis, and P. Bergveld: Sensors and Actuators. Vol. A77 (1999), pp.223-228.
2. C. Yamahata, F. Lacharme, Y. Burri, and M.A.M. Gijs: Sensors and Actuators. Vol. B110 (2005), pp.1-7.
3. A. Olsson, G. Stemme, and E. Stemme: Sensors and Actuators. Vol. A57 (1996), p.137–143.
4. B. Wang, X. Chu, E. Li, and L. Li: Ultrasonics. Vol. 44 (2006), pp.643-646.
5. W.L. Benard, H. Kahn, A.H. Heuer, and M.A. Huff: J. of Microelectromechanical Systems. Vol. 7 (1998), pp.245-251.