The EMF-PF Fill Height Measuring Technique Using Multi-Electrodes in Parallel Connection

Author:

He Long Qing1,Wei Kai Xia1

Affiliation:

1. Nanjing Xiaozhuang College

Abstract

A new sensor of the electromagnetic flowmeter in partially filled pipes (EMF-PF) is presented. With an array of parallel electrodes constituted by multi-electrodes in parallel connection as the measurement electrodes, the measurement electrodes alternate duties, sensing both the flow velocity and fill height signals. On the sensor pipe bottom, one pair of the electrodes is set as the signal injection electrodes that inject the AC signal used to determine liquid fill height. The theoretical analysis and experiment results indicate it feasible for the EMF-PF to employ this fill height measuring technique .

Publisher

Trans Tech Publications, Ltd.

Reference8 articles.

1. Cai Wuchang, Ma Zhongyuan, In: The Electromagnetic flow- meter. China Petroleum & Chemical Press, (2005).

2. Flood,J. Single-Sensor Measurement of Flow in Partially Filled Process Pipes. Sensors Magazine, Sep. (1997).

3. Blake Doney, ABB Automation, Inc. EMF Flow Measure - ment in Partially Filled Pipes. Sensors October, (1999).

4. Zhang Tao, Xu Ling. an. Inverstigation on Measurement of flow in part-filled pipes based on calculation method. Chi- nese journal of scientific instrument. Vol. 17, No. 1, Jan, (1996).

5. O'sullivan V T. Performance of an electromagnetic flowme- ter with six point electrodes. Phys E: Sci Instrum, 1983, Vol. 16, No. 12, pp.1183-1188.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献