Author:
Blamey Peter J.,Martin Lois F.A.,Fiket Hayley J.
Abstract
A new amplification strategy (ADRO™), based on 64 independently operating channels, was compared with a nine-channel wide dynamic range compression strategy (WDRC). Open-platform in-the-ear hearing instruments were configured either with ADRO or the manufacturer's WDRC strategy. Twenty-two subjects with mild to moderate hearing loss took home the ADRO or WDRC hearing aids. After three weeks' acclimatization, the aids were evaluated using monosyllables in quiet at 50 to 65 dB SPL and sentences in eight-talker babble. The acclimatization and evaluation were repeated in the second phase of the balanced reverse-block blind experimental design. The ADRO program showed a statistically significant mean advantage of 7.85% word score (95% confidence interval 3.19% to 12.51%; p = 0.002) and 6.41% phoneme score for the monosyllables in quiet (95% confidence interval 2.03% to 10.79%; p = 0.006). A statistically significant advantage of 7.25% was also found for the ADRO program in background noise (95% confidence interval 1.95% to 12.55%; p = 0.010). The results are consistent with earlier data for listeners with moderate to severe hearing loss.
Cited by
14 articles.
订阅此论文施引文献