Author:
Kalkanis James G.,Krishna Priya,Espinosa Jose A.,Naritoku Dean K.
Abstract
✓ Vagus nerve stimulation for treatment of epilepsy is considered safe; reports of severe complications are rare. The authors report on two developmentally disabled patients who experienced vocal cord paralysis weeks after placement of a vagus nerve stimulator. In both cases, traction injury to the vagus nerve resulting in vocal cord paralysis was caused by rotation of the pulse generator at the subclavicular pocket by the patient. Traumatic vagus nerve injury caused by patients tampering with their device has never been reported and may be analogous to a similar phenomenon reported for cardiac pacemakers in the literature. As the use of vagus nerve stimulation becomes widespread it is important to consider the potential for this adverse event.
Publisher
Journal of Neurosurgery Publishing Group (JNSPG)
Cited by
28 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献