Acute failure of S2-alar-iliac screw pelvic fixation in adult spinal deformity: novel failure mechanism, case series, and review of the literature

Author:

Martin Christopher T.1,Polly David W.12,Holton Kenneth J.1,San Miguel-Ruiz Jose E.1,Albersheim Melissa1,Lender Paul1,Sembrano Jonathan N.1,Hunt Matthew A.2,Jones Kristen E.12

Affiliation:

1. Department of Orthopaedic Surgery, University of Minnesota; and

2. Department of Neurosurgery, University of Minnesota, Minneapolis, Minnesota

Abstract

OBJECTIVE Pelvic fixation with S2-alar-iliac (S2AI) screws is an established technique in adult deformity surgery. The authors’ objective was to report the incidence and risk factors for an underreported acute failure mechanism of S2AI screws. METHODS The authors retrospectively reviewed a consecutive series of ambulatory adults with fusions extending 3 or more levels, and which included S2AI screws. Acute failure of S2AI screws was defined as occurring within 6 months of the index surgery and requiring surgical revision. RESULTS Failure occurred in 6 of 125 patients (5%) and consisted of either slippage of the rods or displacement of the set screws from the S2AI tulip head, with resultant kyphotic fracture. All failures occurred within 6 weeks postoperatively. Revision with a minimum of 4 rods connecting to 4 pelvic fixation points was successful. Two of 3 (66%) patients whose revision had less fixation sustained a second failure. Patients who experienced failure were younger (56.5 years vs 65 years, p = 0.03). The magnitude of surgical correction was higher in the failure cohort (number of levels fused, change in lumbar lordosis, change in T1–pelvic angle, and change in coronal C7 vertical axis, each p < 0.05). In the multivariate analysis, younger patient age and change in lumbar lordosis were independently associated with increased failure risk (p < 0.05 for each). There was a trend toward the presence of a transitional S1–2 disc being a risk factor (OR 8.8, 95% CI 0.93–82.6). Failure incidence was the same across implant manufacturers (p = 0.3). CONCLUSIONS All failures involved large-magnitude correction and resulted from stresses that exceeded the failure loads of the set plugs in the S2AI tulip, with resultant rod displacement and kyphotic fractures. Patients with large corrections may benefit from 4 total S2AI screws at the time of the index surgery, particularly if a transitional segment is present. Salvage with a minimum of 4 rods and 4 pelvic fixation points can be successful.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Reference40 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3