Robotic external ventricular drain placement for acute neurosurgical care in low-resource settings: feasibility considerations and a prototype design

Author:

Smith Alexander D.1,Teague Alexander J.1,Naik Anant1,Janbahan Mika1,Smith Emily J.1,Krist David T.1,Parupalli Sindhu1,Teal Kevin12,Hassaneen Wael12

Affiliation:

1. Carle Illinois College of Medicine, University of Illinois at Urbana-Champaign, Urbana; and

2. Department of Neurological Surgery, Neuroscience Institute, Carle Foundation Hospital, Urbana, Illinois

Abstract

OBJECTIVE Emergency neurosurgical care in lower-middle-income countries faces pronounced shortages in neurosurgical personnel and infrastructure. In instances of traumatic brain injury (TBI), hydrocephalus, and subarachnoid hemorrhage, the timely placement of external ventricular drains (EVDs) strongly dictates prognosis and can provide necessary stabilization before transfer to a higher-level center of care that has access to neurosurgery. Accordingly, the authors have developed an inexpensive and portable robotic navigation tool to allow surgeons who do not have explicit neurosurgical training to place EVDs. In this article, the authors aimed to highlight income disparities in neurosurgical care, evaluate access to CT imaging around the world, and introduce a novel, inexpensive robotic navigation tool for EVD placement. METHODS By combining the worldwide distribution of neurosurgeons, CT scanners, and gross domestic product with the incidence of TBI, meningitis, and hydrocephalus, the authors identified regions and countries where development of an inexpensive, passive robotic navigation system would be most beneficial and feasible. A prototype of the robotic navigation system was constructed using encoders, 3D-printed components, machined parts, and a printed circuit board. RESULTS Global analysis showed Montenegro, Antigua and Barbuda, and Seychelles to be primary candidates for implementation and feasibility testing of the novel robotic navigation system. To validate the feasibility of the system for further development, its performance was analyzed through an accuracy study resulting in accuracy and repeatability within 1.53 ± 2.50 mm (mean ± 2 × SD, 95% CI). CONCLUSIONS By considering regions of the world that have a shortage of neurosurgeons and a high incidence of EVD placement, the authors were able to provide an analysis of where to prioritize the development of a robotic navigation system. Subsequently, a proof-of-principle prototype has been provided, with sufficient accuracy to target the ventricles for EVD placement.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Neurology (clinical),General Medicine,Surgery

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3