Does robot-assisted navigation influence pedicle screw selection and accuracy in minimally invasive spine surgery?

Author:

Shafi Karim A.1,Pompeu Yuri A.1,Vaishnav Avani S.1,Mai Eric1,Sivaganesan Ahilan1,Shahi Pratyush1,Qureshi Sheeraz A.1

Affiliation:

1. Hospital for Special Surgery, New York, New York

Abstract

OBJECTIVE The accuracy of percutaneous pedicle screw placement has increased with the advent of robotic and surgical navigation technologies. However, the effect of robotic intraoperative screw size and trajectory templating remains unclear. The purpose of this study was to compare pedicle screw sizes and accuracy of placement using robotic navigation (RN) versus skin-based intraoperative navigation (ION) alone in minimally invasive lumbar fusion procedures. METHODS A retrospective cohort study was conducted using a single-institution registry of spine procedures performed over a 4-year period. Patients who underwent 1- or 2-level primary or revision minimally invasive surgery (MIS)–transforaminal lumbar interbody fusion (TLIF) with pedicle screw placement, via either robotic assistance or surgical navigation alone, were included. Demographic, surgical, and radiographic data were collected. Pedicle screw type, quantity, length, diameter, and the presence of endplate breach or facet joint violation were assessed. Statistical analysis using the Student t-test and chi-square test was performed to evaluate the differences in pedicle screw sizes and the accuracy of placement between both groups. RESULTS Overall, 222 patients were included, of whom 92 underwent RN and 130 underwent ION MIS-TLIF. A total of 403 and 534 pedicle screws were placed with RN and ION, respectively. The mean screw diameters were 7.25 ± 0.81 mm and 6.72 ± 0.49 mm (p < 0.001) for the RN and ION groups, respectively. The mean screw length was 48.4 ± 4.48 mm in the RN group and 45.6 ± 3.46 mm in the ION group (p < 0.001). The rates of “ideal” pedicle screws in the RN and ION groups were comparable at 88.5% and 88.4% (p = 0.969), respectively. The overall screw placement was also similar. The RN cohort had 63.7% screws rated as good and 31.4% as acceptable, while 66.1% of ION-placed screws had good placement and 28.7% had acceptable placement (p = 0.661 and p = 0.595, respectively). There was a significant reduction in high-grade breaches in the RN group (0%, n = 0) compared with the ION group (1.2%, n = 17, p = 0.05). CONCLUSIONS The results of this study suggest that robotic assistance allows for placement of screws with greater screw diameter and length compared with surgical navigation alone, although with similarly high accuracy. These findings have implied that robotic platforms may allow for safe placement of the “optimal screw,” maximizing construct stability and, thus, the ability to obtain a successful fusion.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Neurology (clinical),General Medicine,Surgery

Reference47 articles.

1. Facet joint violation during percutaneous pedicle screw placement: a comparison of two techniques;Tannous O,2017

2. Accuracy of current techniques for placement of pedicle screws in the spine: a comprehensive systematic review and meta-analysis of 51,161 screws. World Neurosurg;Perdomo-Pantoja A,2019

3. Structural characteristics of the pedicle and its role in screw stability;Hirano T,1997

4. Impact of screw diameter on pedicle screw fatigue strength-a biomechanical evaluation. World Neurosurg;Viezens L,2021

5. Pathology and pathogenesis of lumbar spondylosis and stenosis;Kirkaldy-Willis WH,1978

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3