Setup accuracy of spine radiosurgery using cone beam computed tomography image guidance in patients with spinal implants

Author:

Gerszten Peter C.12,Monaco Edward A.1,Quader Mubina23,Novotny Josef23,Kim Jong Oh23,Flickinger John C.23,Huq M. Saiful23

Affiliation:

1. 1Departments of Neurological Surgery and

2. 2Radiation Oncology, University of Pittsburgh Medical Center; and

3. 3Pittsburgh Cancer Institute, Pittsburgh, Pennsylvania

Abstract

Object Cone beam computed tomography (CBCT) image guidance technology has been adopted for use in spine radiosurgery. There is concern regarding the ability to safely and accurately perform spine radiosurgery without the use of implanted fiducials for image guidance in postsurgical cases in which titanium instrumentation and/or methylmethacrylate (MMA) has been implanted. In this study the authors prospectively evaluated the accuracy of the patient setup for spine radiosurgery by using CBCT image guidance in the context of orthopedic hardware at the site of disease. Methods The positioning deviations of 31 single-fraction spine radiosurgery treatments in patients with spinal implants were prospectively evaluated using the Elekta Synergy S 6-MV linear accelerator with a beam modulator and CBCT image guidance combined with a robotic couch that allows positioning correction in 3 translational and 3 rotational directions. To measure patient movement, 3 quality-assurance CBCT studies were performed and recorded: before, halfway through, and after radiosurgical treatment. The positioning data and fused images of planning CTs and CBCTs from the treatments were analyzed to determine intrafractional patient movements. From each of 3 CBCTs, 3 translational and 3 rotational coordinates were obtained. Results The prescribed dose to the gross tumor volume for the cohort was 12–18 Gy (mean 14 Gy) utilizing 9–14 coplanar intensity-modulated radiation therapy (IMRT) beams (mean 10 beams). At the halfway point of the radiosurgery, the translational variations and standard deviations were 0.6 ± 0.6, 0.4 ± 0.4, and 0.5 ± 0.5 mm in the lateral (X), longitudinal (Y), and anteroposterior (Z) directions, respectively. The magnitude of the 3D vector (X,Y,Z) was 1.1 ± 0.7 mm. Similarly, the variations immediately after treatment were 0.5 ± 0.3, 0.4 ± 0.4, and 0.5 ± 0.6 mm along the X, Y, and Z directions, respectively. The 3D vector was 1.0 ± 0.6 mm. The mean rotational angles were 0.3 ± 0.4, 0.5 ± 0.6, and 0.3 ± 0.4° along yaw, roll, and pitch, respectively, at the halfway point and 0.3 ± 0.4, 0.6 ± 0.6, and 0.4 ± 0.5° immediately after treatment. Conclusions Cone beam CT image guidance used for patient setup for spine radiosurgery was highly accurate despite the presence of spinal instrumentation and/or MMA at the level of the target volume. The presence of such spinal implants does not preclude safe treatment via spine radiosurgery in these patients.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3