The singular value decomposition (SVD) based clutter filter is commonly applied to beamformed signals for the visualization of echo signals from flowing blood cells. In this paper, the SVD-based clutter filter is applied to signals directly acquired from ultrasonic elements before beamforming to be compared with the conventional strategy by evaluating contrast and standard deviation (SD) in the filtered images. As a result, the contrast was improved from 10.7 ± 3.6 dB to 18.3 ± 4.6 dB, and the SD was slightly improved from 3.78 ± 0.69 dB to 3.07 ± 0.74 dB in measurement of a right jugular vein.