Nosocomial Transmission of Imipenem-Resistant Pseudomonas Aeruginosa Following Bronchoscopy Associated With Improper Connection to the STERIS SYSTEM 1 Processor

Author:

Sorin Mircea,Segal-Maurer Sorana,Mariano Noriel,Urban Carl,Combest Anna,Rahal James J.

Abstract

AbstractObjective:To assess nosocomial transmission of imipenem-resistant Pseudomonas aeruginosa (IRPA) following bronchoscopy during August through October 1998.Design:Traditional and molecular epidemiological investigation of a case series.Setting:University-affiliated community hospital.Patients:18 patients with IRPA bronchial-wash isolates.Interventions:We reviewed clinical data, performed environmental cultures and molecular analysis of all IRPA isolates, and observed disinfection of bronchoscopes.Results:Of 18 patients who had IRPA isolated from bronchoscopic or postbronchoscopic specimens, 13 underwent bronchoscopy for possible malignancy or undiagnosed pulmonary infiltrates. Following bronchoscopy, 3 patients continued to have IRPA isolated from sputum and demonstrated clinical evidence of infection requiring specific antimicrobial therapy. The remaining 15 patients had no further IRPA isolated and remained clinically well 3 months following bronchoscopy. Pulsed-field gel electrophoresis revealed that all strains except one were >95% related. STERIS SYSTEM 1 had been implemented in July 1998 as an automatic endoscope reprocessor (AER) for all endoscopes and bronchoscopes. Inspection of bronchoscope sterilization cycles revealed incorrect connectors joining the bronchoscope suction channel to the STERIS SYSTEM 1 processor, obstructing peracetic acid flow through the bronchoscope lumen. No malfunction warning was received, and spore strips remained negative.Conclusions:The similarity of diverse connectors and limited training by the manufacturer regarding AER for bronchoscopes were the two factors responsible for the outbreak. Appropriate connections were implemented, and there was no further bronchoscope contamination. We suggest active surveillance of all bronchoscopy specimen cultures, standardization of connectors of various scopes and automated processors, and systematic education of staff by manufacturers with periodic on-site observation.

Publisher

Cambridge University Press (CUP)

Subject

Infectious Diseases,Microbiology (medical),Epidemiology

Cited by 72 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3