Affiliation:
1. Department of Pharmaceutical Analysis, GITAM Institute of Pharmacy, GITAM (Deemed to be University), Visakhapatnam-530045, India
Abstract
The dosage form iotrim (gramicidin, neomycin and triamcinolone) is a combination of two antibiotics
(gramicidin and neomycin) and a steroid (triamcinolone). The antibiotics work by killing the bacteria
that cause infections. The steroid blocks the action of chemical messengers (prostaglandins) that make
the affected area red, swollen and itchy. Consequently, there was still a need to develop a simple, less
time consuming and economical method for the simultaneous determination of gramicidin, neomycin
and triamcinolone acetonide. The current work is an effort to develop a fast and reproducible LC-MS
technique for the simultaneous estimation of gramicidin, neomycin and triamcinolone acetonide. The
objective of the present procedure was to validate and develop a precise and accurate liquid
chromatography-mass spectrometry (LC-MS) technique for the simultaneous quantification of
gramicidin, neomycin and triamcinolone acetonide. Gramicidin, neomycin and triamcinolone acetonide
were monitored on Shimadzu-8045 mass spectrometer equipped with electro spray ionization interface.
The retention times of gramicidin, neomycin and triamcinolone acetonide were found at 9.145 min,
7.273 min and 2.435 min, respectively. The limit of detection (LOD) results for gramicidin, neomycin
and triamcinolone acetonide were observed to be 0.15, 1.5 and 0.6 μg/mL, respectively while the limit
of quantification (LOQ) results were observed to be 0.5, 5, 2 μg/mL concentration, respectively. The
linear range for gramicidin, neomycin and triamcinolone acetonide were found in the concentration
ranges from 1.25-7.5 μg/mL, 12.5-75 μg/mL and 5-30 μg/mL with regression coefficient of 0.9991,
0.9996, 0.9999, respectively. Accuracy values for gramicidin, neomycin and triamcinolone acetonide
were found to be in the range of 98.64%, 99.4%, 99.5% respectively. The % RSD for six replicates in
precision was less than 2%. According to ICH Q2(R1) recommendations, this method was successfully
tested with LC-MS to confirm the chemical structures of newly produced degradation products of
triamcinolone acetonide and neomycin. The developed process was validated efficaciously as per ICH
guidelines.
Publisher
Asian Journal of Chemistry