An in vitro method for comparing biocompatibility of materials for extracorporeal circulation

Author:

Peek Giles J1,Scott Richard2,Killer Hilliary M3,Firmin Richard K3

Affiliation:

1. Division of Cardiac Surgery, Department of Surgery, Glenfield Hospital, University of Leicester, University Hospitals of Leicester NHS Trust, Leicester, UK,

2. Glenfield Hospital, University Hospitals of Leicester NHS Trust, Leicester, UK

3. Heartlink ECMO Centre, Department of Cardiothoracic Surgery, Glenfield Hospital, University Hospitals of Leicester NHS Trust, Leicester, UK

Abstract

We measured the response of fresh heparinized human blood to recirculation through circuits made of LVA (Portex Industries, Hythe, Kent, UK), SRT (Rehau UK, Langley, Slough, UK) and Tygon® S-65-HL (Norton Performance Plastics, Corby, Northants, UK), as control. Circuit construction: 1/2 in. tubing, heat exchanger (Dideco D-720P), Stockert roller pump, just underoccluded, Cincinnati Sub Zero heater, circuit volume of 500 ml. Flow 3.45 l/min, 37°C. Samples: at 10 min, 1, 2, 4 and 6 h. n= 5 in each group; 2/5 SRT experiments were stopped at 45 and 60 min due to overpressurization. Results: Baseline activated clotting time (ACT) of 300 s, increasing in all groups as fibrinogen fell to zero with SRT and LVA. Minimum fibrinogen was 1 g/l for Tygon. Absolute thrombocytopenia occurred (SRT and LVA 60 min and Tygon 240 min). International normalized ratio (INR) in both the SRT and LVA circuits increased, but mean increase for Tygon (0.56) was smaller than the other two materials. Plasma free haemoglobin increased in all three materials; the increase was greater in the LVA circuits compared to the control. C5b9 levels increased equally in all groups. Lactoferrin levels rose equally in all groups to a maximum at 150 min. The neutrophil counts fell, mirroring the lactoferrin. The total white cell counts also fell in all groups; in the LVA circuits, the fall was significantly lower than in the control. Rapid disappearance of platelets and fibrinogen from the blood in the SRT and LVA circuits excludes them both from extracorporeal use. Paradoxically, SRT caused the least complement activation of the three materials. This method can be used to compare biocompatibility.

Publisher

SAGE Publications

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3