Test and Development of a Specialized Pipeline for Ventilator Calibration

Author:

Liu Junming,Zheng Wenxiu,Wang Ziyan

Abstract

A specialized Y-type pipeline is developed to replace medical disposable Y-type pipelines for ventilator calibration and to improve accuracy, providing a reference for improvement. According to the calibration specifications of ventilators, tests are performed on specialized pipelines and more to compare their data. Taking the tidal volume 400-mL detection point as an example, the calibration errors made by medical disposable pipelines and specialized pipelines are 6.2% and −0.8%, respectively, and the errors at other detection points are roughly the same. The accuracy of the specialized Y-type pipeline has significantly improved, reducing the impact of the pipeline on tidal volume calibration from more than 6% to less than 1%. The use of specialized pipelines can significantly improve the accuracy and qualification rate of calibration, reducing the consumption of disposable Y-type pipelines and thereby significantly reducing costs and increasing efficiency.

Publisher

Begell House

Subject

Biomedical Engineering

Reference26 articles.

1. Global health concern stirred by emerging viral infections;Luo G.X.;J Med Virol,2020

2. Nucleic acid detection of the SARS-CoV-2;Zhang Y.Z.;Acta Metrol Sinica,2020

3. Early transmission dynamics in Wuhan, China, of novel coronavirus-infected pneumonia;Li Q.;N Engl J Med,2020

4. Preparation of SARS-CoV-2 genomic RNA reference materials;Wang D.;Acta Metrol Sinica,2021

5. First case of 2019 novel coronavirus in the United States;Holshue M.L.;N Engl J Med,2020

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3