Abstract
This work presents the development of a non-enzymatic electrochemical sensor for creatinine in a neutral medium using a cobalt-gold bimetallic nanoparticles modified platinum electrode. The voltammetric detection of creatinine in a neutral phosphate buffer was based on the formation of a soluble cobalt-creatinine complex. The sensor exhibited good selectivity and a detection limit (S/N = 3) of 2.25 mM with two linear ranges from 6.4 to 83.2 mM. The sensitivity of the sensor was 0.621 and 1.135 μA mM−1 cm−2 at lower (6.4–51.2 mM) and higher (51.2–83.2 mM) detection ranges, respectively. The sensor performance was validated using urine samples and creatinine spiked urine samples.
Funder
Department of Biotechnology (DBT), Government of India
Publisher
The Electrochemical Society