Abstract
Real-time detection of L-lactate is crucial in monitoring tissue oxygenation and organ metabolism in surgery patients during the perioperative period. Traditional commercial L-lactate detection techniques using bulky and expensive instruments are time-consuming, hindering timely feedback in the operating room. Herein, utilizing the modified screen-printed carbon electrode, a portable and user-friendly smartphone-based electrochemical biosensing system, integrated with a miniaturized potentiostat via wireless transmission, was successfully constructed for bedside detection of L-lactate. It has a wide sensing range from 0.05–10 mM and a low limit of detection (9.1 μM). Bland–Altman analysis shows an acceptable agreement with the values of a mean difference (0.114 ± 0.3482) between results obtained with the commercial blood gas analysis instrument and the developed portable system. Accordingly, the developed portable system exhibits a great potential in point-of-care testing owing to its portability and rapid response in L-lactate detection during perioperative monitoring.
Funder
National Natural Science Foundation of China
Publisher
The Electrochemical Society
Subject
Materials Chemistry,Electrochemistry,Surfaces, Coatings and Films,Condensed Matter Physics,Renewable Energy, Sustainability and the Environment,Electronic, Optical and Magnetic Materials
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献