Abstract
Introduction: Aortic angulation (AA), defined as the angle between the aortic annulus plane and the horizontal plane, may result in failed prosthesis positioning. The effect of AA on the procedural and short-term outcomes with the portico valves for transcatheter aortic valve implantation (TAVI) has not been fully investigated. The present study aimed to evaluate the impact of AA on device success and early outcomes of TAVI using a selfexpanding portico valve.
Patients and Methods: Preoperative computed tomography scans of 121 consecutive patients treated with the portico valve were analyzed. TAVI device success and outcomes were determined according to VARC-3 definitions. Patients were divided into two groups based on mean AA.
Results: The mean AA was 48.7 ± 8.9°. There were no differences in technical success (92.2 vs 89.5%, p= 0.604), device success (81.2% vs 77.2%, p= 0.582), and early safety endpoints (68.8% vs 61.4%, p= 0.397) between the AA≤ 48° and AA>48° groups. The frequency and severity of paravalvular aortic regurgitation (PAR) was statistically higher in patients with AA> 48° (p= 0.028). Moreover, an increased AA was also associated with valve malposition (12.3% vs. 1.6%, p= 0.018), prolonged procedure time (85 ± 26 vs. 75 ± 20 minutes, p= 0.028), and greater Δimplantation depth (2.2 ± 0.5 vs. 0.8 ± 0.1 mm; p<0.001).
Conclusion: Despite comparable device success and early outcomes rates, increased AA is associated with higher rates of PAR and valve malposition, with the self-expanding portico valve.
Publisher
Kosuyolu Heart Journal, Health Sciences University
Reference27 articles.
1. Smith CR, Leon MB, Mack MJ, Miller DC, Moses JW, Svensson LG, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med 2011;364:2187-98. [Crossref]
2. Thourani VH, Kodali S, Makkar RR, Herrmann HC, Williams M, Babaliaros V, et al. Transcatheter aortic valve replacement versus surgical valve replacement in intermediate-risk patients: A propensity score analysis. Lancet 2016;387:2218-25. [Crossref]
3. Mack MJ, Leon MB, Thourani VH, Makkar R, Kodali SK, Russo M, et al. Transcatheter aortic-valve replacement with a balloon-expandable valve in low-risk patients. N Engl J Med 2019;380:1695-705. [Crossref]
4. Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk patients. N Engl J Med 2019;380:1706-15. [Crossref]
5. Van Mieghem NM, Popma JJ, Deeb GM, Yakubov SJ, Serruys PW, Windecker S, et al. Complete 2-year results confirm bayesian analysis of the SURTAVI trial. JACC Cardiovasc Interv 2020;13:323-31. [Crossref]