Variations in Determining Actual Orientations of Segmented Deep Brain Stimulation Leads Using the DiODe Algorithm: A Retrospective Study Across Different Lead Designs and Medical Institutions

Author:

Henry Kaylee R.ORCID,Miulli Milina MichelleORCID,Nuzov Noa B.ORCID,Nolt Mark J.,Rosenow Joshua M.ORCID,Elahi BehzadORCID,Pilitsis Julie,Golestanirad Laleh

Abstract

<b><i>Introduction:</i></b> Directional deep brain stimulation (DBS) leads have become widely used in the past decade. Understanding the asymmetric stimulation provided by directional leads requires precise knowledge of the exact orientation of the lead in respect to its anatomical target. Recently, the DiODe algorithm was developed to automatically determine the orientation angle of leads from the artifact on postoperative computed tomography (CT) images. However, manual DiODe results are user-dependent. This study analyzed the extent of lead rotation as well as the user agreement of DiODe calculations across the two most common DBS systems, namely, Boston Scientific’s <i>Vercise</i> and Abbott’s <i>Infinity</i>, and two independent medical institutions. <b><i>Methods:</i></b> Data from 104 patients who underwent an anterior-facing unilateral/bilateral directional DBS implantation at either Northwestern Memorial Hospital (NMH) or Albany Medical Center (AMC) were retrospectively analyzed. Actual orientations of the implanted leads were independently calculated by three individual users using the DiODe algorithm in Lead-DBS and patients’ postoperative CT images. The deviation from the intended orientation and user agreement were assessed. <b><i>Results:</i></b> All leads significantly deviated from the intended 0° orientation (<i>p</i> &lt; 0.001), regardless of DBS lead design (<i>p</i> &lt; 0.05) or institution (<i>p</i> &lt; 0.05). However, the Boston Scientific leads showed an implantation bias toward the left at both institutions (<i>p</i> = 0.014 at NMH, <i>p</i> = 0.029 at AMC). A difference of 10° between at least two users occurred in 28% (NMH) and 39% (AMC) of all Boston Scientific and 76% (NMH) and 53% (AMC) of all Abbott leads. <b><i>Conclusion:</i></b> Our results show that there is a significant lead rotation from the intended surgical orientation across both DBS systems and both medical institutions; however, a bias toward a single direction was only seen in the Boston Scientific leads. Additionally, these results raise questions into the user error that occurs when manually refining the orientation angles calculated with DiODe.

Publisher

S. Karger AG

Subject

Neurology (clinical),Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3