Abstract
Background: Implantable pulse generators (IPGs) store energy and deliver electrical impulses for deep brain stimulation (DBS) to treat neurological and psychiatric disorders. IPGs have evolved over time to meet the demands of expanding clinical indications and more nuanced therapeutic approaches. Objectives: The aim of this study was to examine the workflow of the first 4-lead IPG for DBS in patients with complex disease. Method: The engineering capabilities, clinical use cases, and surgical technique are described in a cohort of 12 patients with epilepsy, essential tremor, Parkinson’s disease, mixed tremor, and Tourette’s syndrome with comorbid obsessive-compulsive disorder between July 2021 and July 2022. Results: This system is a rechargeable 32-channel, 4-port system with independent current control that can be connected to 8 contact linear or directionally segmented leads. The system is ideal for patients with mixed disease or those with multiple severe symptoms amenable to >2 lead implantations. A multidisciplinary team including neurologists, radiologists, and neurosurgeons is necessary to safely plan the procedure. There were no serious intraoperative or postoperative adverse events. One patient required revision surgery for bowstringing. Conclusions: This new 4-lead IPG represents an important new tool for DBS surgery with the ability to expand lead implantation paradigms for patients with complex disease.
Subject
Neurology (clinical),Surgery
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献