Abstract
The validation of the electrothermal spectrometry method for Arsenic and lead determination in water samples
guarantees the quality of analytical data. The study was carried out at AQLAB. A theoretical and technical
investigation was applied to ensure the reliability and accuracy of the analytical method. Parameters such as
INEN standards, Eurachem international standards, AQLAB internal documents and environmental standards
were used to validate the process. Several readings were performed on different samples to make calibration
curves for As and Pb, evaluate the method's linearity, and obtain information on the slope, intercept and determination coefficient. Statistical calculations were used to determine the limit of detection and quantification,
repeatability, reproducibility, trueness and uncertainty. The results show that the hypotheses were fulfilled,
calibration curves with determination coefficient higher than 0.995, repeatability standard deviation lower
than 16%, calculated F < critical F (4.96), calculated t student < theoretical (2.23), trueness was between
100.29-110.18 and 99.64-107.92%, uncertainty was 20% range 0.005 to 0.10 mg/L and 15% range 0.01-
0.20mg/L, limit of detection was 0.001 and 0.005 mg/L and limit of quantification was 0.01 and 0.02mg/L,
respectively. Therefore, the validation method was robust and accurate.
Keywords: water samples, analytical data, reliable results, Arsenic, Lead, validation
Reference48 articles.
1. 1. Hill SJ, Fisher AS. Atomic Absorption, Methods and Instrumentation. En: Lindon JC, Tranter GE, Koppenaal DW, editores. Encyclopedia of Spectroscopy and Spectrometry (Third Edition) [Internet]. Oxford:
2. Bionatura Journal journal 2024, 10.21931/BJ/2024.01.01.8 12
3. Clinical Biotec, Universidad Católica del Oriente (UCO) and Universidad Nacional Autónoma de Honduras (UNAH)
4. Academic Press; 2017 [citado 24 de julio de 2023]. p. 37-43. Disponible en: https://www.sciencedirect.com/science/article/pii/B9780128032244000996
5. 2. Gallegos W, Vega M, Noriega P. Espectroscopía de absorción atómica con llama y su aplicación para la