Achieved Versus Intended Pulse Oximeter Saturation in Infants Born Less Than 28 Weeks' Gestation: The AVIOx Study

Author:

Hagadorn James I.1,Furey Anne M.2,Nghiem Tuyet-Hang3,Schmid Christopher H.4,Phelps Dale L.5,Pillers De-Ann M.6,Cole Cynthia H.7,

Affiliation:

1. Division of Neonatology, Connecticut Children's Medical Center, Hartford, Connecticut

2. Division of Newborn Medicine, Floating Hospital for Children/Tufts-New England Medical Center, Boston, Massachusetts

3. Division of Neonatology, Children's Hospital Boston, Boston, Massachusetts

4. Institute for Clinical Research and Health Policy Studies, Tufts-New England Medical Center, Boston, Massachusetts

5. Departments of Pediatrics and Ophthalmology, University of Rochester School of Medicine and Dentistry, Rochester, New York

6. Departments of Pediatrics, Ophthalmology, and Molecular & Medical Genetics, Doernbecher Children's Hospital, Casey Eye Institute, Oregon Health & Science University, Portland, Oregon

7. Department of Neonatology, Beth Israel-Deaconess Medical Center, Boston, Massachusetts

Abstract

OBJECTIVE. The objective of this study was to document pulse oximeter saturation levels achieved in the first 4 weeks of life in infants who were born at <28 weeks' gestation, compared with the levels that were targeted by local policy, and examine factors that are associated with compliance with the target range. METHODS. Infants who were <28 weeks' gestation and ≤96 hours of age were enrolled in a prospective, multicenter cohort study. Oximetry data were collected with masked signal-extraction oximeters for a 72-hour period in each of the first 4 weeks of life. Data were compared with the pulse oximeter saturation target range prescribed by local institutional policy. Factors that were associated with intended range compliance were identified with hierarchical modeling. RESULTS. Fourteen centers from 3 countries enrolled 84 infants with mean ± SD birth weight of 863 ± 208 g and gestational age of 26 ± 1.4 weeks. Oxygen saturation policy limits ranged between 83% and 92% for lower limits and 92% and 98% for upper limits. For infants who received respiratory support, median pulse oximeter saturation level achieved was 95%. Center-specific medial levels were within the intended range at 12 centers. Centers maintained infants within their intended range 16% to 64% of the time but were above range 20% to 73% of the time. In hierarchical modeling, wider target ranges, higher target range upper limits, presence of a policy of setting oximeter alarms close to the target range limits, and lower gestational age were associated with improved target range compliance. CONCLUSIONS. Success with maintaining the intended pulse oximeter saturation range varied substantially among centers, among patients within centers, and for individual patients over time. Most noncompliance was above the intended range. Methods for improving compliance and the effect of improved compliance on neonatal outcomes require additional research.

Publisher

American Academy of Pediatrics (AAP)

Subject

Pediatrics, Perinatology and Child Health

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3