Clinical Performance of an In-Line Point-of-Care Monitor in Neonates

Author:

Widness John A.1,Kulhavy Jeff C.2,Johnson Karen J.1,Cress Gretchen A.1,Kromer Irma J.1,Acarregui Michael J.1,Feld Ronald D.2

Affiliation:

1. From the Department of Pediatrics, Children's Hospital of Iowa, and the

2. Department of Pathology, University of Iowa, Iowa City, Iowa.

Abstract

Objective. To evaluate the bias, precision, and blood loss characteristics of an ex vivo in-line point-of-care testing blood gas and electrolyte monitor designed for use in critically ill newborn infants. Study Design. Study participants included consecutive neonates with an umbilical artery catheter (UAC) in use for clinical laboratory testing. The in-line monitor (VIA LVM Blood Gas and Chemistry Monitoring System, VIA Medical, San Diego, CA) was directly connected to the participant's UAC and the monitor's determinations of pH, Pco2, Po2, sodium, potassium, and hematocrit (Hct) were compared with those simultaneously drawn and measured with a standard bench top laboratory instrument (Radiometer 625 ABL; Radiometer America, Inc, Westlake, OH). The bias (the mean difference from the reference method) and precision (1 standard deviation of the mean difference) performance criteria of the in-line monitor were derived using standard laboratory procedures. Results. Sixteen neonates monitored for a total of 37 days had a total of 229 paired blood samples available for comparison by the 2 methods. Bias and precision performance characteristics of the in-line monitor were similar to reports of other point-of-care devices (ie, pH: −.003 ± .024; Pco2: .35 ± 2.84 mm Hg; Po2: .39 ± 7.30 mm Hg; sodium: .52 ± 2.34 mmol/L; potassium: .17 ± .18 mmol/L; and Hct: .61 ± 2.80%). The range of values observed for each parameter included much of the range anticipated among critically ill neonates (ie, pH: 7.15–7.65; Pco2: 25–75 mm Hg; Po2: 25–275 mm Hg; sodium: 127–150 mmol/L; potassium: 2.6–5.5 mmol/L; and Hct: 32%–60%). Mean blood loss (± standard deviation) per sample with the in-line monitor was approximately one-tenth that of the reference method: 24 ± 7 μL versus 250 μL, respectively. There was no evidence of hemolysis and no patient related safety issues were identified with use of the in-line monitor. Conclusions. Repeated laboratory testing of critically ill neonates using an ex vivo in-line monitor designed for use in neonates provides reliable laboratory results. The blood loss and hemolysis data obtained suggests that this monitoring device offers potential for reducing neonatal blood loss—and possibly transfusion needs—during the first weeks of life. Before this promising technology can be routinely recommended for care of critically ill neonates, greater practical experience in a variety of clinical settings is needed.

Publisher

American Academy of Pediatrics (AAP)

Subject

Pediatrics, Perinatology, and Child Health

Cited by 42 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1.

Anemia in Pediatric Critical Care

;International Journal of Clinical Transfusion Medicine;2020-10

2. Monitoring with In Vivo Electrochemical Sensors: Navigating the Complexities of Blood and Tissue Reactivity;Sensors;2020-06-02

3. Patterns of phlebotomy blood loss and transfusions in extremely low birth weight infants;Journal of Perinatology;2019-10-03

4. Hematological diagnostics;Point-of-Care Testing;2018

5. Hämatologische Diagnostik;POCT - Patientennahe Labordiagnostik;2017

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3