Assessment the Accuracy of Densitometry Measurements Using DMA PP2 Phantom

Author:

Petraikin A. V.1,Smoliarchuk M. Ya.1,Petryaykin F. A.2,Nisovtsova L. A.1,Artyukova Z. R.3,Sergunova K. A.1,Akhmad E. S.1,Semenov D. S.1,Vladzymyrsky A. V.1,Morozov S. P.1

Affiliation:

1. Practical and Clinical Research Center of Diagnostics and Telemedicine Technologies

2. Lomonosov Moscow State University

3. Sechenov First Moscow State Medical University (Sechenov University)

Abstract

Purpose of the study — to assess the accuracy of dual energy x-ray absorptiometry (DXA ) for measurements of mineral bone density, bone mineral content, area of selected spine zone of examination as well as impact of subcutaneous fat layer and correction of auto-segmenting of the spine on the mentioned parameters. Material and Methods. The study was performed on iDXA scanner using the designed phantom DMA PP2 of the lumber spine with inlays to simulate subcutaneous fat (SF). To ensure correct assessment of measurements (precision and accuracy) the authors performed fivefold repeated scanning. Two modifications of the phantom were used, with and without SF inlays, as well as two methods for selection of spine range for examination – automatic and correction of autosegmentation. Results. Scanning of the phantom without SF inlays demonstrated a systematic understated values of bone mineral density (BMD) and bone mineral content (BMC) along the full measured interval: mean relative error of BMD for L1-L4 interval was 10.62% with automatic segmentation and 7.43% — with  correction of autosegmentation. The least accuracy for BMD and BMC (1.53% and 0.90%, respectively) was observed during SF simulation and with correction of auto-segmentation of the spine. Analysis of variation coefficient for area of examined vertebrae, BMC and BMD demonstrated rather high precision of measurements, namely for BMD without SF in the L1-L4 interval amounted to 1.00% (auto-segmentation) and 0.56% (correction). Variation coefficient for scanning including SF inlays in the interval L1-L4 was 1.00% (auto-segmentation) and 0.68% (correction). Conclusion. The lowest level of accuracy was observed with the SFL object; in this case, the variation coefficient did not exceed 1% for all BMD interval. The mean value of the BMC accuracy also did not exceed 1% with the optimal scan parameters. The study proved the effectiveness of “RSK PK2” phantom when estimating the accuracy of BMD and BMC on iDXA scanner.

Publisher

ECO-Vector LLC

Reference23 articles.

1. Mikhailov E.E., Benevolenskaya L.I. [Epidemiology of osteoporosis and fractures]. In: [Osteoporosis Guide]. Moscow: BINOM. Laboratoriya znanii; 2003. р. 10-55. (In Russian).

2. Lesnyak O., Ershova O., Belova K., Gladkova E., Sinitsina O., Ganert O. et al. Epidemiology of fracture in the Russian Federation and the development of a FRAX model. Arch Osteoporos. 2012;7:67-73.

3. 2019 ISCD Official Positions - Adult. Available from: https://www.iscd.org/official-positions/2019-iscdofficial-positions-adult/.

4. Avrunin A.S., Tikhilov R.M., Shubniakov I.I. [Dynamic assessment of bone tissue osteocytic remodeling using a non-invasive method]. [Morphology]. 2009;135(2):66-73. (in Russian).

5. Mel’nichenko G.A., Belaya Zh.E., Rozhinskaya L.Ya., Toroptsova N.V., Alekseeva L.I., Biryukova E.V. et al. Russian federal clinical guidelines on the diagnostics, treatment, and prevention of osteoporosis. [Problems of Endocrinology]. 2017;(6):392-426. (In Russian). doi: 10.14341/probl2017636392-426.

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3