A Second-Generation Blackbody System for the Calibration and Verification of Seagoing Infrared Radiometers

Author:

Donlon Craig J.1,Wimmer W.2,Robinson I.2,Fisher G.2,Ferlet M.3,Nightingale T.3,Bras B.1

Affiliation:

1. European Space Agency, Noordwijk, Netherlands

2. National Oceanography Centre, University of Southampton, Southampton, United Kingdom

3. Space Science and Technology Department, Rutherford Appleton Laboratory, Didcot, United Kingdom

Abstract

AbstractQuasi-operational shipborne radiometers provide a fiducial reference measurement (FRM) for satellite validation of satellite sea surface skin temperature (SSTskin) retrievals. External reference blackbodies are required to verify the performance and to quantify the accuracy of the radiometer calibration system. They provide a link in an unbroken chain of comparisons between the shipborne radiometer and a traceable reference standard. A second-generation water bath blackbody reference radiance source has been developed for this purpose. The second generation Concerted Action for the Study of the Ocean Thermal Skin (CASOTS-II) blackbody has a 110-mm-diameter aperture cylinder-cone geometry coated with NEXTEL suede 3103 paint. Interchangeable aperture stops reduce the cavity aperture diameter and minimize stray radiation. Monte Carlo modeling techniques show the effective emissivity of the cavity to be >0.9999 (aperture < 30 mm). The cavity is immersed in a water bath that is vigorously stirred using a pump that slowly heats the water bath at a mean rate of ~0.6 K h−1. The temperature of the water bath is measured using a thermometer traceable to the International System of Units (SI) standards. The worst-case radiance temperature of the CASOTS-II blackbody system is traceable to the SI with an uncertainty of 58 mK (millikelvin). When operating under typical laboratory conditions using an aperture of 40 mm, the uncertainty is 16 mK. An intercomparison with the U.K. National Physical Laboratory Absolute Measurements of Blackbody Emitted Radiance (AMBER) reference radiometer found no significant differences within 75 mK (110-mm aperture) or 50 mK (40-mm aperture), which is the combined uncertainty of the comparison and the reference standard for SI traceability of ISAR radiometer SSTskin records used for satellite SST validation. Applications of the CASOTS-II blackbody to monitor the calibration of shipborne radiometers are described and measurement protocols are proposed.

Publisher

American Meteorological Society

Subject

Atmospheric Science,Ocean Engineering

Reference42 articles.

1. The Miami2001 infrared radiometer calibration and intercomparison. Part II: Shipboard results;Barton;J. Atmos. Oceanic Technol.,2004

2. Emissivity of a cylindrical black-body cavity with a re-entrant cone end face;Berry;J. Phys.,1981

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3