Intercenter Agreement in Reading Doppler Embolic Signals

Author:

Markus Hugh S.1,Ackerstaff Rob1,Babikian Viken1,Bladin Chris1,Droste Dirk1,Grosset Donald1,Levi Chris1,Russell David1,Siebler Mario1,Tegeler Charles1

Affiliation:

1. From the Department of Clinical Neuroscience (H.S.M.), King’s College School of Medicine and Dentistry and the Institute of Psychiatry, London, UK; the Department of Neurophysiology (R.A.), St Antonius Ziekenhuis, Niuewegein, Netherlands; the Department of Neurology (V.B.), Boston (Mass) University School of Medicine; the Department of Neurology (C.B., C.L.), Austin and Repatriation Medical Centre, Melbourne, Australia; the Department of Neurology (D.D.), University of Munster (Germany); the...

Abstract

Background and Purpose Different frequencies of asymptomatic Doppler embolic signals have been reported in studies. There has been concern that different criteria for identification may account for some of this variation. A previous reproducibility study between two centers found good agreement, but no studies among large numbers of centers have been performed. We performed an international reproducibility study among nine centers, each of which had published recent studies of embolic signal detection in peer-reviewed journals. Methods Each center performed blinded analysis of a taped audio Doppler signal composed of transcranial Doppler middle cerebral artery recordings from 6 patients with symptomatic carotid artery stenosis. The exact time of any embolic signal was recorded. Six centers also measured the intensity increase of any embolic signals detected. Interobserver agreement was determined by a method based on the proportion of specific agreement. Results Seven centers reported between 39 and 55 signals, but one center reported 142 embolic signals. The probability of agreement between observers was .678, which rose to .791 when the data from the highest reporting center were excluded. Introducing a decibel threshold resulted in a significant increase in the probability of agreement; a decibel threshold of >7 dB resulted in a probability of agreement of .902. Intensity measurements made by different centers were usually highly correlated, but this was not always the case, and 3 of the 15 correlations were not significant. The absolute values of the intensities measured varied between centers by as much as 40%. Conclusions Although most centers report similar numbers of embolic signals, some use less specific criteria and report more events. The use of a decibel threshold improves reproducibility. However, intensity thresholds developed by one center cannot be directly transferred without validation to another center; differing methods of measurement are being used, and this results in different intensity values for the same embolic signals, even when the same equipment is used.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Clinical Neurology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3