Replicating Patient-Specific Severe Aortic Valve Stenosis With Functional 3D Modeling

Author:

Maragiannis Dimitrios1,Jackson Matthew S.1,Igo Stephen R.1,Schutt Robert C.1,Connell Patrick1,Grande-Allen Jane1,Barker Colin M.1,Chang Su Min1,Reardon Michael J.1,Zoghbi William A.1,Little Stephen H.1

Affiliation:

1. From the Houston Methodist DeBakey Heart and Vascular Center, Houston, TX (D.M., M.S.J., S.R.I., R.C.S., C.M.B., S.M.C., M.J.R., W.A.Z., S.H.L.); and Department of Bioengineering, Rice University, Houston, TX (P.C., J.G.-A.).

Abstract

Background— 3D stereolithographic printing can be used to convert high-resolution computed tomography images into life-size physical models. We sought to apply 3D printing technologies to develop patient-specific models of the anatomic and functional characteristics of severe aortic valve stenosis. Methods and Results— Eight patient-specific models of severe aortic stenosis (6 tricuspid and 2 bicuspid) were created using dual-material fused 3D printing. Tissue types were identified and segmented from clinical computed tomography image data. A rigid material was used for printing calcific regions, and a rubber-like material was used for soft tissue structures of the outflow tract, aortic root, and noncalcified valve cusps. Each model was evaluated for its geometric valve orifice area, echocardiographic image quality, and aortic stenosis severity by Doppler and Gorlin methods under 7 different in vitro stroke volume conditions. Fused multimaterial 3D printed models replicated the focal calcific structures of aortic stenosis. Doppler-derived measures of peak and mean transvalvular gradient correlated well with reference standard pressure catheters across a range of flow conditions ( r =0.988 and r =0.978 respectively, P <0.001). Aortic valve orifice area by Gorlin and Doppler methods correlated well ( r =0.985, P <0.001). Calculated aortic valve area increased a small amount for both methods with increasing flow ( P =0.002). Conclusions— By combing the technologies of high-spatial resolution computed tomography, computer-aided design software, and fused dual-material 3D printing, we demonstrate that patient-specific models can replicate both the anatomic and functional properties of severe degenerative aortic valve stenosis.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging

Cited by 140 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3