Validation of Noninvasive Measurement of Cardiac Output Using Inert Gas Rebreathing in a Cohort of Patients With Heart Failure and Reduced Ejection Fraction

Author:

Hassan Mohamed1,Wagdy Kerolos1,Kharabish Ahmed1,Selwanos Peter Philip1,Nabil Ahmed1,Elguindy Ahmed1,ElFaramawy Amr1,Elmahdy Mahmoud F.1,Mahmoud Hani1,Yacoub Magdi H.1

Affiliation:

1. From the Cardiology Department (M.H., A.E., M.F.E.) and Radiology Department (A.K.), Cairo University, Egypt; Cardiology Department (M.H., K.W., P.P.S., A.E., H.M.), Radiology Department (A.K.), Biomedical Engineering Department (A.N.), and Cardiothoracic Department (M.H.Y.), Aswan Heart Centre, Egypt; and Imperial College, London, United Kingdom (M.H.Y.).

Abstract

Background— Cardiac output (CO) is a key indicator of cardiac function in patients with heart failure. No completely accurate method is available for measuring CO in all patients. The objective of this study was to validate CO measurement using the inert gas rebreathing (IGR) method against other noninvasive and invasive methods of CO quantification in a cohort of patients with heart failure and reduced ejection fraction. Methods and Results— The study included 97 patients with heart failure and reduced ejection fraction (age 42±15.5 years; 64 patients (65.9%) had idiopathic dilated cardiomyopathy and 21 patients (21.6%) had ischemic heart disease). Median left ventricle ejection fraction was 24% (10%–36%). Patients with atrial fibrillation were excluded. CO was measured using 4 methods (IGR, cardiac magnetic resonance imaging, cardiac catheterization, and echocardiography) and indexed to body surface area (cardiac index [CI]). All studies were performed within 48 hours. Median CI measured by IGR was 1.75, by cardiac magnetic resonance imaging was 1.82, by cardiac catheterization was 1.65, and by echo was 1.7 L·min −1 ·m −2 . There were significant modest linear correlations between IGR-derived CI and cardiac magnetic resonance imaging–derived CI ( r =0.7; P <0.001), as well as cardiac catheterization–derived CI ( r =0.6; P <0.001). Using Bland–Altman analysis, the agreement between the IGR method and the other methods was as good as the agreement between any 2 other methods with each other. Conclusions— The IGR method is a simple, accurate, and reproducible noninvasive method for quantification of CO in patients with advanced heart failure. The prognostic value of this simple measurement needs to be studied prospectively.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3