Abstract

Background— Sprint Fidelis (Fidelis) implantable cardioverter-defibrillator leads are prone to fractures that have caused adverse events, primarily inappropriate shocks, and a few reported deaths. More than 100 000 patients have Fidelis leads. No independent multicenter long-term performance information exists for this lead, and single-center studies suggest that certain patients are at increased risk for Fidelis failure. Our aim was to assess the longevity of Fidelis leads and to evaluate clinical variables and adverse events associated with Fidelis failure. The results were compared with like data for Quattro Secure (Quattro) implantable cardioverter-defibrillator leads implanted at our centers. Methods and Results— This 3-center study included adults ≥18 years of age who received Fidelis or Quattro leads for the prevention of sudden cardiac death. From November 2001 to January 2009,1023 Fidelis and 1668 Quattro leads were implanted and followed up. The failure rate for Fidelis leads was 2.81%/y compared with 0.43%/y for Quattro leads ( P <0.0001). No deaths or injuries occurred as a result of lead failure, but 42% of fractures caused inappropriate shocks. The survival of Fidelis leads at 4 years was 87.0% (95% confidence interval, 83.6 to 90.1) compared with 98.7% (95% confidence interval, 97.9 to 99.4) for Quattro leads ( P <0.0001). Multivariate predictors of Fidelis failure were younger age (hazard ratio, 0.98; 95% confidence interval, 0.96 to 0.99), female gender (hazard ratio, 0.61; 95% confidence interval, 0.40 to 1.00), and cardiac disease ( P =0.041). Conclusions— Compared with Quattro leads, the survival of Fidelis leads continues to decline, and Fidelis failure is notably higher in younger patients, women, individuals with hypertrophic cardiomyopathy, and patients with arrhythmogenic right ventricular dysplasia or channelopathies. These findings have significant implications for the management of patients who have Fidelis leads, and they demonstrate the importance of weighing clinical variables in assessments of implantable cardioverter-defibrillator lead performance.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 112 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3