BACKGROUND
Necropsy studies demonstrate that coronary artery disease (CAD) is frequently complex and eccentric. However, angiography provides only a silhouette of the vessel lumen. Intravascular ultrasound is a new tomographic imaging method for evaluation of coronary dimensions and wall morphology. Few data exist regarding intravascular ultrasound in patients with CAD, and no data exist for subjects with normal coronaries.
METHODS AND RESULTS
We used a multielement 5.5F, 20-MHz ultrasound catheter to examine eight normal subjects and 43 patients with CAD. We assessed the safety of coronary ultrasound and the effect of vessel eccentricity on comparison of minimum luminal diameter by angiography and ultrasound. Normal and atherosclerotic wall morphology and stenosis severity were also evaluated by intravascular ultrasound. The only untoward effect was transient coronary spasm in five patients. At 33 sites in normal subjects, the lumen was nearly circular, yielding a close correlation between angiographic and ultrasonic minimum diameter (r = 0.92). At 90 sites in patients with CAD, ultrasound demonstrated a concentric cross section; correlation was also close (r = 0.93). However, at 72 eccentric sites, correlation was not as close (r = 0.77). For 41 stenoses, correlation between angiography and ultrasound for area reduction was moderate (r = 0.63). In normal subjects, wall morphology revealed a thin (0.30 mm or less) intimal leading edge and subadjacent sonolucent zone (0.20 mm or less). Patients with CAD exhibited increased thickness and echogenicity of the leading edge, thickened sonolucent zones, and/or attenuation of ultrasound transmission.
CONCLUSIONS
These data establish that intravascular ultrasound is feasible and safe and yields luminal measurements that correlate generally with angiography. Differences between angiographic and ultrasound measures of lumen size in eccentric vessels probably reflect the dissimilar perspectives of tomographic and silhouette imaging techniques. Intravascular ultrasound provides detailed images of normal and abnormal wall morphology not previously possible in vivo.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Cited by
605 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献