Efficacy of a single-lead unipolar transvenous defibrillator compared with a system employing an additional coronary sinus electrode. A prospective, randomized study.

Author:

Kudenchuk P J1,Bardy G H1,Dolack G L1,Poole J E1,Mehra R1,Johnson G1

Affiliation:

1. Department of Medicine, University of Washington, Seattle 98195.

Abstract

BACKGROUND Recent development of a prototype single-lead unipolar transvenous defibrillator offers the possibility of device implantation with the ease of a permanent pacemaker. Lowering defibrillation energy requirements would allow for a further reduction in defibrillator generator size and enhance the feasibility of pacemaker-like placement. However, if achieving a lower defibrillation energy requires placing additional intracardiac leads, the potential advantage of a smaller generator may be offset by the disadvantages of a more complex lead system. The purpose of this study was to compare defibrillation energy requirements of a single-lead unipolar defibrillator with a three-electrode system employing an additional lead in the coronary sinus. METHODS AND RESULTS Testing of a single-lead unipolar biphasic pulse defibrillation system versus a three-electrode system with an additional coronary sinus lead was performed in prospective, randomized fashion in 15 patients with a history of ventricular tachycardia or fibrillation. Ventricular fibrillation was induced with alternating current, and defibrillation threshold was measured by a pulse given 10 seconds after arrhythmia induction. The mean defibrillation threshold stored energy and mean leading edge voltage did not significantly differ between the two systems (11.3 +/- 5.9 J versus 9.9 +/- 5.2 J and 418 +/- 118 V versus 390 +/- 112 V, respectively; P > .4). Using either defibrillation system, all patients were successfully defibrillated by < 24 J and over half of patients by < 10 J. CONCLUSIONS A unipolar transvenous biphasic defibrillation system is an effective means of treating ventricular fibrillation. The added complexity of additional leads is not offset by any significant improvement in defibrillation efficacy or energy requirements. Given the simplicity and effectiveness of a single-lead system coupled with a small generator, placement of defibrillation systems may now approach the ease of pacemaker implantation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3