Temperature-guided radiofrequency catheter ablation with very large distal electrodes.

Author:

Langberg J J1,Gallagher M1,Strickberger S A1,Amirana O1

Affiliation:

1. Department of the Internal Medicine, University of Michigan Medical Center, Ann Arbor 48109-0022.

Abstract

BACKGROUND Previous studies have shown that the size of lesions produced by radiofrequency catheter ablation correlates with the temperature and surface area of the electrode-tissue interface. The purpose of the present study was to compare the effects of ablation using very large distal electrodes (8F, 8 and 12 mm long) with those made by a conventional radiofrequency ablation catheter (distal electrode 8F, 4 mm long). METHODS AND RESULTS Each catheter had a thermistor in the tip of the distal electrode. Radiofrequency energy (500 kHz) was supplied by a generator that continuously monitored temperature and produced up to 100 W. In 10 dogs, each of the three ablation catheters were introduced percutaneously and positioned under fluoroscopic guidance at disparate left ventricular endocardial sites. Radiofrequency power output was titrated to achieve a temperature of 80 degrees C for 60 seconds at each ablation site. The power required to produce a steady-state temperature of 80 degrees C was directly proportional to electrode size (15 +/- 7, 46 +/- 15, and 62 +/- 32 W using the 4-mm-, 8-mm-, and 12-mm-long electrodes, respectively). Lesions produced by the 8-mm electrode were nearly twice as deep (11 +/- 2.4 versus 6 +/- 1.2 mm, P < .001) and four times as large (905 +/- 410 versus 210 +/- 100 mm3, P < .001) as those made with a conventional 4-mm electrode. Lesions produced by the 12-mm electrode were intermediate in size (depth, 8 +/- 1.2 mm; volume, 465 +/- 225 mm3) and sometimes were associated with charring and crater formation. Ablation with the larger electrodes caused a drop in arterial pressure and more ventricular ectopy than ablation using a 4-mm distal electrode. CONCLUSIONS Thermistor-equipped elongated ablation electrodes coupled to high-power outputs can reproducibly produce lesions approximately 1 cm in diameter. This system may prove useful for ablation of ventricular tachycardias in patients with coronary artery disease.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 107 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Autonomous three-dimensional sensor-assisted hybrid force/position control for tendon-driven catheters;The European Physical Journal Plus;2023-06-06

2. Radiofrequency ablation;Principles and Technologies for Electromagnetic Energy Based Therapies;2022

3. New Polish prototypes of 4- and 8-mm-tip nonirrigated radiofrequency ablation catheters: an in vitro study;Kardiologia Polska;2020-11-25

4. Can an Expanding Lattice Electrode Catheter Expand Our Success in Catheter Ablation?;Circulation: Arrhythmia and Electrophysiology;2019-04

5. Irrigated and Cooled-Tip Radiofrequency Catheter Ablation;Catheter Ablation of Cardiac Arrhythmias;2019

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3