Efficacy of automatic multimodal device therapy for ventricular tachyarrhythmias as delivered by a new implantable pacing cardioverter-defibrillator. Results of a European multicenter study of 102 implants.

Author:

Fromer M1,Brachmann J1,Block M1,Siebels J1,Hoffmann E1,Almendral J1,Ohm O J1,den Dulk K1,Coumel P1,Camm A J1

Affiliation:

1. Division de Cardiologie, CHUV, Lausanne, Switzerland.

Abstract

BACKGROUND Third-generation implantable cardioverter-defibrillators are devices designed to treat ventricular tachycardia (VT) and ventricular fibrillation (VF) by means of overdrive pacing, cardioversion, or defibrillation. So far, the efficacy of tiered therapy has been documented only in small series. Therefore, a European multicenter clinical evaluation study of a new tachyarrhythmia control device, the Medtronic PCD pacer-cardioverter-defibrillator with epicardial patch-lead configuration, was undertaken. METHODS AND RESULTS We report on 102 patients (mean age, 55 +/- 13 years) from 11 European centers. PCD devices implanted between May 1989 and February 1991 were included. The patients suffered from hemodynamically significant ventricular tachyarrhythmias not suppressed by antiarrhythmic drug therapy and unrelated to acute myocardial infarction; one patient had nonsustained VT and severely depressed left ventricular function. Seventy patients had coronary artery disease with old myocardial infarctions, 23 had cardiomyopathies of various etiologies, and nine patients had no detectable heart disease. Mean ejection fraction was 36 +/- 14% (range, 10-76%). Mean intraoperative defibrillation threshold (51 patients) was 10.6 +/- 5.1 J (range, 2-18 J). The documented follow-up ranged from 1 to 21 months (mean, 9.4 +/- 5.8 months), or 79.9 cumulative patient-years. Perioperative mortality was 3.9%. The actuarial survival rate at 12 months was 91%. One sudden arrhythmic death occurred. Sixty patients (58%) received device therapy. Seventeen patients had therapies only for "VF" episodes, 16 patients only for VT, and 28 patients for VT and "VF" episodes. Based on device memory data, 1,235 spontaneous VT episodes were detected and treated in 43 patients. Twelve hundred four of these VT episodes received painless initial antitachycardia pacing therapy, restoring sinus rhythm in 91%. The 108 ongoing episodes received 209 multiple therapeutic attempts. Eighty-five additional overdrive pacing therapies restored sinus rhythm in 30%. Initial ineffective antitachycardia pacing therapies received 51 cardioversion pulses. The success rate was 61%. Seventy-three additional cardioversion pulses were delivered to backup ineffective pacing therapy as well as ineffective secondary cardioversion pulses. Their success rate was only 40%. Two hundred eighty-six spontaneous episodes were detected in 44 patients as "VF." Overall defibrillation efficacy was 97.6%. CONCLUSIONS The implanted device nearly eliminates sudden arrhythmic death in patients with documented, potentially fatal ventricular tachyarrhythmias. Automatic tiered therapy is highly effective to restore sinus rhythm, provided that an integrated two-zone tachycardia detection algorithm is used, assigning lower tachycardia rates to overdrive pacing and/or cardioversion and higher tachycardia rates to defibrillation. In general, spontaneous VTs can be terminated by automatic overdrive pacing, and painful or disturbing countershock therapies are not required to terminate the majority of spontaneous VT episodes.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Cited by 116 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3