Inverse Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Author:

Nakagawa Hiroshi1,Wittkampf Fred H. M.1,Yamanashi William S.1,Pitha Jan V.1,Imai Shinobu1,Campbell Barclay1,Arruda Mauricio1,Lazzara Ralph1,Jackman Warren M.1

Affiliation:

1. From the Cardiovascular Section/Department of Medicine, University of Oklahoma Health Sciences Center and the Department of Veterans Affairs Medical Center, Oklahoma City; and the Department of Cardiology, Heart Lung Institute, University Hospital Utrecht, Netherlands (F.H.M.W.).

Abstract

Background —Clinical efficacy has driven the use of larger electrodes (7F, length ≥4 mm) for radiofrequency ablation, which reduces electrogram resolution and causes variability in tissue contact depending on electrode orientation. With active cooling, ablation electrode size may be reduced. The purpose of this study was to examine the effect of electrode length on tissue temperature and lesion size with saline irrigation used for active cooling. Methods and Results —In 11 anesthetized dogs, the thigh muscle was exposed and bathed with heparinized canine blood. A 7F ablation catheter with a 2- or 5-mm irrigated tip electrode was positioned perpendicular or parallel to the thigh muscle. Radiofrequency current was delivered at constant voltage (50 V) for 30 seconds during saline irrigation (20 mL/min) to 148 sites. Tissue temperature at depths of 3.5 and 7 mm and lesion size were measured. In the perpendicular electrode-tissue orientation, radiofrequency applications at 50 V with the 2-mm electrode compared with the 5-mm electrode resulted in lower power at 50 V (26 versus 36 W) but higher tissue temperatures, larger lesion depth (8.0 versus 5.4 mm), and greater diameter (12.4 mm versus 8.4 mm). Also, in the parallel orientation, overall power was lower with the 2-mm electrode (25 versus 33 W), but tissue temperatures were higher and lesions were deeper (7.3 versus 6.9 mm). Lesion diameter was similar (11.1 versus 11.3 mm) for both electrodes. Conclusions —The smaller electrode resulted in transmission of a greater fraction of the radiofrequency power to the tissue and resulted in higher tissue temperature, larger lesions, and lower dependency of lesion size on the electrode orientation.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3