Multicenter Comparison of Truncated Biphasic Shocks and Standard Damped Sine Wave Monophasic Shocks for Transthoracic Ventricular Defibrillation

Author:

Bardy Gust H.1,Marchlinski Francis E.1,Sharma Arjun D.1,Worley Seth J.1,Luceri Richard M.1,Yee Raymond1,Halperin Blair D.1,Fellows Christopher L.1,Ahern Thomas S.1,Chilson Donald A.1,Packer Douglas L.1,Wilber David J.1,Mattioni Thomas A.1,Reddy Ramakota1,Kronmal Richard A.1,Lazzara Ralph1

Affiliation:

1. the Department of Medicine, University of Washington (Seattle).

Abstract

Background The most important factor for improving out-of-hospital ventricular fibrillation survival rates is early defibrillation. This can be achieved if small, lightweight, inexpensive automatic external defibrillators are widely disseminated. Because automatic external defibrillator size and cost are directly affected by defibrillation waveform shape and because of the favorable experience with truncated biphasic waveforms in implantable cardioverter-defibrillators, we compared the efficacy of a truncated biphasic waveform with that of a standard damped sine monophasic waveform for transthoracic defibrillation. Methods and Results The principal goal of this multicenter, prospective, randomized, blinded study was to compare the first-shock transthoracic defibrillation efficacy of a 130-J truncated biphasic waveform with that of a standard 200-J monophasic damped sine wave pulse using anterior thoracic pads in the course of implantable cardioverter-defibrillator testing. Pad-pad ECGs were also examined after transthoracic defibrillation. After the elimination of data for 24 patients who did not meet all protocol criteria, the results from 294 patients were analyzed. The 130-J truncated biphasic pulse and the 200-J damped sine wave monophasic pulse resulted in first-shock efficacy rates of 86% and 86%, respectively ( P =.97). ST-segment levels measured 10 seconds after the shock in 151 patients in sinus rhythm were −0.26±1.58 and −1.86±1.93 mm for the 130- and 200-J shocks, respectively ( P <.0001). Conclusions We found that 130-J biphasic truncated transthoracic shocks defibrillate as well as the 200-J monophasic damped sine wave shocks that are traditionally used in standard transthoracic defibrillators and result in fewer ECG abnormalities after the shock.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference64 articles.

1. Poole JP Bardy GH. Sudden cardiac death. In: Zipes DP Jalife J eds. Cardiac Electrophysiology: From Cell to Bedside 2nd ed. Philadelphia Pa: WB Saunders; 1995:812-832.

2. Out-of-hospital cardiac arrest: Factors associated with survival

3. Cardiac arrest and resuscitation: A tale of 29 cities

4. Outcome of Out-of-Hospital Cardiac Arrest in New York City

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3