1. From the Department of Medicine (P.B.A.), Cardiovascular Disease and Physiology, University of Oklahoma Health Sciences Center, Oklahoma City, Okla; the Carlyle Fraser Heart Center/Division of Cardiology (A.L.S.), Emory University School of Medicine, Atlanta, Ga; Ohio State University Heart Center (W.T.A.), Columbus, Ohio; and Medtronic, Inc (K.J.K., R.W.S., A.S., M.M.R.), Minneapolis, Minn.

Abstract

Background— Heart rate variability (HRV) as an indirect autonomic assessment provides prognostic information when measured over short time periods in patients with heart failure. Long-term continuous HRV can be measured from an implantable device, but the clinical value of these measurements is unknown. Methods and Results— A total of 397 patients with New York Heart Association class III or IV heart failure were studied. Of these, 370 patients had information from their implanted cardiac resynchronization device for mortality risk stratification, and 288 patients had information for measured parameters (ie, HRV, night heart rate, and patient activity) and clinical event analyses. Continuous HRV was measured as the standard deviation of 5-minute median atrial-atrial intervals (SDAAM) sensed by the device. SDAAM <50 ms when averaged over 4 weeks was associated with increased mortality risk (hazard ratio 3.20, P =0.02) and SDAAM were persistently lower over the entire follow-up period in patients who required hospitalization or died. SDAAM decreased a median of 16 days before hospitalization and returned to baseline after treatment. Automated detection of decreases in SDAAM was 70% sensitive in detecting cardiovascular hospitalization, with 2.4 false-positives per patient-year of follow-up. Conclusions— This study demonstrates that SDAAM continuously measured from an implanted cardiac resynchronization device is lower in patients at high mortality and hospitalization risk. SDAAM declines as patient status decompensates. Continuous long-term SDAAM may be a useful tool in the clinical management of patients with chronic heart failure.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3