Clinical and Hemodynamic Results of Intraaortic Balloon Pumping and Surgery for Cardiogenic Shock

Author:

DUNKMAN W. BRUCE1,LEINBACH ROBERT C.1,BUCKLEY MORTIMER J.1,MUNDTH ELDRED D.1,KANTROWITZ ARTHUR R.1,AUSTEN W. GERALD1,SANDERS CHARLES A.1

Affiliation:

1. From the Medical and Surgical Services of the Massachusetts General Hospital and the Departments of Medicine and Surgery, Harvard Medical School, Boston, Massachusetts.

Abstract

The AVCO balloon pump has been employed in treating 40 patients with cardiogenic shock from acute myocardial infarction (CS-MI). All patients were given a trial of medical therapy with hemodynamic monitoring. The time from the development of shock to institution of intraaortic balloon pumping (IABP) was less than 24 hours in all but nine patients. Prior to IABP the mean hemodynamic values were: cardiac index (CI) 1.7 liters/min/m 2 ; mean arterial pressure (MAP) 66 mm Hg; pulmonary artery wedge pressure (PAW) 22 mm Hg. After 24-48 hours of IABP the CI and MAP had increased 0.8 liters/min/m 2 and 8 mm Hg, respectively, and the PAW had decreased 4.8 mm Hg. During IABP the shock syndrome was reversed in 31 patients. Four of 25 patients treated with IABP alone survived to be discharged, but two have died from subsequent infarctions. Because of the persistent high mortality, 15 patients judged unable to survive off IABP have undergone emergency surgical procedures with IABP continuing during preoperative angiography and postoperatively. Six were long-term survivors. It is concluded: (1) IABP is a safe, effective means of supporting the circulation in CS-MI; (2) IABP alone will improve survival in some patients; (3) IABP can provide circulatory support during angiography and the perioperative period in patients requiring revascularization for survival; and (4) some patients with CS-MI have myocardial necrosis too extensive to permit survival without permanent circulatory assistance or total cardiac replacement.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference23 articles.

1. Diastolic balloon pumping (with carbon dioxide) in the aorta—A mechanical assistance to the failing circulation

2. Mechanical Intraaortic Cardiac Assistance in Cardiogenic Shock

3. com.atypon.pdfplus.internal.model.plusxml.impl.AuthorGroup@25d9f98b : Hemodynamic evaluation of intra-aortic balloon pumping in man. Circulation 41 (suppl II): II-130 1970

4. Clinical management ancillary to phase-shift balloon pumping in cardiogenic shock

5. Clinical experience with the unidirectional dual-chambered intra-aortic balloon assist;Circulation,1971

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3