Abstract
Background
—To distinguish prolonged episodes of atrial fibrillation (AF) that require cardioversion from self-terminating episodes that do not, an atrial implantable cardioverter-defibrillator (ICD) must be able to detect AF continuously for extended periods. The ICD should discriminate between atrial tachycardia/flutter (AT), which may be terminated by antitachycardia pacing, and AF, which requires cardioversion.
Methods and Results
—We studied 80 patients with AT/AF and ventricular arrhythmias who were treated with a new atrial/dual-chamber ICD. During a follow-up period lasting 6±2 months, we validated spontaneous, device-defined AT/AF episodes by stored electrograms in all patients. In 58 patients, we performed 80 Holter recordings with telemetered atrial electrograms, both to validate the continuous detection of AT/AF and to determine the sensitivity of the detection of AT/AF. Detection was appropriate in 98% of 132 AF episodes and 88% of 190 AT episodes (98% of 128 AT episodes with an atrial cycle length <300 ms). Intermittent sensing of far-field R waves during sinus tachycardia caused 27 inappropriate AT/AF detections; these detections lasted 2.6±2.0 minutes. AT/AF was detected continuously in 27 of 28 patients who had spontaneous episodes of AT/AF (96%). The device memory recorded 90 appropriate AT/AF episodes lasting >1 hour, for a total of 2697 hours of continuous detection of AT/AF. During Holter monitoring, the sensitivity of the detection of AT/AF (116 hours) was 100%; the specificity of the detection of non-AT/AF rhythms (1290 hours) was 99.99%. Of 166 appropriate episodes detected as AT, 45% were terminated by antitachycardia pacing.
Conclusions
—A new ICD detects AT/AF accurately and continuously. Therapy may be programmed for long-duration AT/AF, with a low risk of underdetection. Discrimination of AT from AF permits successful pacing therapy for a significant fraction of AT.
Publisher
Ovid Technologies (Wolters Kluwer Health)
Subject
Physiology (medical),Cardiology and Cardiovascular Medicine
Reference12 articles.
1. Gillberg J Brown M Stanton M Hammill S Olson W. Clinical testing of a dual chamber atrial tachyarrhythmia detection algorithm. Comput Cardiol. 1996;57–60.
2. Brown M Gillberg J Stanton M Hammill S Olson W. Acute human testing of a dual chamber ventricular tachyarrhythmia detection algorithm at Mayo Clinic. Comput Cardiol . 1996;61–64.
3. Electrical Proarrhythmia: Induction of Inappropriate Atrial Therapies due to Far-Field R Wave Oversensing in a New Dual Chamber Defibrillator
4. Mode Switch Despite Undersensing of Atrial Fibrillation in ODD Pacing
5. Is Mode Switching Beneficial? A Randomized Study in Patients With Paroxysmal Atrial Tachyarrhythmias
Cited by
155 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献