Randomized Evaluation of Polytetrafluoroethylene-Covered Stent in Saphenous Vein Grafts

Author:

Stankovic Goran1,Colombo Antonio1,Presbitero Patrizia1,van den Branden Frank1,Inglese Luigi1,Cernigliaro Carmelo1,Niccoli Luigi1,Bartorelli Antonio L.1,Rubartelli Paolo1,Reifart Nicholaus1,Heyndrickx Guy R.1,Saunamäki Kari1,Morice Marie Claude1,Sgura Fabio A.1,Di Mario Carlo1

Affiliation:

1. From Columbus Hospital, Milan, Italy (G.S., A.C., C.D.M.); Istituto Clinico Humanitas (P.P.), Rozzano, Italy; Middelheim Ziekenhuis (F.v.d.B.), Antwerpen, Belgium; Ospedale Clinicizzato (L.I.), San Donato, Milan, Italy; Azienda Ospedaliera (C.C.), Novara, Italy; Ospedale Civile (L.N.), Brescia, Italy; Centro Cardiologico Monzino IRCCS (A.L.B.), University of Milan, Milan, Italy; Ospedale San Martino (P.R.), Genova, Italy; Kardiologisches Institut Main-Taunus (N.R.), Bad Soden, Germany;...

Abstract

Background— Treatment of lesions located in saphenous vein grafts (SVGs) is associated with increased procedural risk and a high rate of restenosis. Methods and Results— We conducted a randomized, multicenter trial to evaluate the usefulness of a polytetrafluoroethylene (PTFE)-covered stent compared with a bare stainless steel (SS) stent for prevention of restenosis and major adverse cardiac events (MACE) in patients undergoing SVG treatment. The primary end point was angiographic restenosis at 6 months. Secondary end points were 30-day and 6-month MACE rates, defined as the cumulative of death, myocardial infarction (MI), and target lesion revascularization. Between September 1999 and January 2002, 301 patients with SVG lesions were randomized to either the PTFE-covered JoStent coronary stent graft (PTFE group, n=156) or the SS JoFlex stent (control group, n=145). Angiographic and procedural success rates were similar between the 2 groups (97.4% versus 97.9% and 87.3% versus 93.8%, respectively). The incidence of 30-day MACE was higher in the PTFE group (10.9% versus 4.1%, P =0.047) and was mainly attributed to MI (10.3% versus 3.4%, P =0.037). The primary end point, the restenosis rate at 6-month follow-up, was similar between the 2 groups (24.2% versus 24.8%, P =0.237). Although the 6-month non–Q-wave MI rate was higher in the PTFE group (12.8% versus 4.1%, P =0.013), the cumulative MACE rate was not different (23.1% versus 15.9%, P =0.153). Conclusions— The study did not demonstrate a difference in restenosis rate and 6-month clinical outcome between the PTFE-covered stent and the SS stent for treatment of SVG lesions. However, a higher incidence of nonfatal myocardial infarctions was found in patients treated with the PTFE-covered stent.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3