Affiliation:
1. State Key Laboratory of Power Transmission Equipment and System Security and New Technology, Chongqing University, Chongqing, 400044, China
Abstract
To improve the image resolution of a portable real-time ultrasound imaging device, this paper proposes a beamformer with composite apodization windows based on Chebyshev polynomial. By introducing the Chebyshev window, the level of sidelobe can be flexibly controlled. The value of weighting changes according to the detection depth and ultrasound attenuation such that it can attain better image resolution than traditional beamformers. The simulations and experiments indicate that the proposed method can flexibly narrow the mainlobe width while keeping the performance of the suppressing artifacts. The full width at half-maximum of this method is 14.1%–27.8% and 4.2%–25.0% narrower than those of traditional apodization methods in the field II simulations and Geabr_0 experiment, respectively.
Funder
NSFC Grant
Chongqing special key project for technological innovation and application development
Publisher
Acoustical Society of America (ASA)
Subject
Acoustics and Ultrasonics,Arts and Humanities (miscellaneous)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献