Evaluation of the AutoMicrobic system for identification of yeasts

Author:

Hasyn J J,Buckley H R

Abstract

The AutoMicrobic system (Vitek Systems, Hazelwood, Mo.) for identification of yeasts was evaluated by testing 352 freshly isolated, germ tube-negative yeasts from clinical laboratories. Of these isolates, 84.9% were correctly identified. The system correctly identified 86.5% of the Candida isolates and 74.3% of the Cryptococcus isolates.

Publisher

American Society for Microbiology

Subject

Microbiology (medical)

Reference19 articles.

1. Microbiologic aspects in patients with leukemia;Bodey G. P.;Hum. Pathol.,1974

2. Infections in cancer patients on a protected environment-prophylactic antibiotic program;Bodey G. P.;Am. J. Med.,1975

3. Evaluation of commercial systems for the identification of clinical yeast isolates;Bowman P. I.;J. Clin. Microbiol.,1976

4. Evaluation of the modified API 20C system for identification of clinically important yeasts;Buesching W. J.;J. Clin. Microbiol.,1979

5. Clinical evaluation of the Uni-Yeast-Tek system fer rapid presumptive identification of medically important yeasts;Cooper B. H.;J. Clin. Microbiol.,1978