Affiliation:
1. National Jewish Medical and Research Center, Denver, Colorado,1 and
2. Missouri State Laboratory, Mt. Vernon, Missouri2
Abstract
ABSTRACT
Two manual liquid medium systems, the Mycobacteria Growth Indicator Tube (MGIT) and MB Redox tube systems, were evaluated in comparison to the radiometric BACTEC-460 semiautomated system for recovery of
Mycobacterium tuberculosis
from sputum specimens. The highest level of recovery, from a total of 77 culture-positive specimens, occurred with the BACTEC-460 system (92.2%), followed by the MB Redox tube (80.5%) and the MGIT (63.6%) systems. The shortest time to detection was observed also among the cultures in BACTEC-460: a mean of 12 days to a growth index (GI) of 10 and 15 days to a GI of 500. The mean times for the other systems were 16 days for the MB Redox tube system and 17.4 days for the MGIT system. The proportion of cultures grown after more than 3 weeks of incubation was only 2.8 or 8.4% in BACTEC-460 (for a GI of 10 or 500) but 17.7% in MB Redox and 22.5% in MGIT. Despite these differences in comparison to the BACTEC-460 system and some differences between the MGIT and MB Redox tube systems, either of the two manual liquid medium systems presents a reasonable alternative to the BACTEC-460 system, especially for laboratories with a limited workload, and a valuable element in the laboratory protocol, in conjunction with solid media, for obtaining rapid detection of growth from about 80% of culture-positive specimens and for better overall recovery of
M. tuberculosis
.
Publisher
American Society for Microbiology
Cited by
31 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献