Corrosion of a Fassier-Duval Telescopic Nail Causing Pain and Osteolysis

Author:

Ngo David1ORCID,Todd Maya1ORCID,Accadbled Franck1ORCID,Foster Bruce1ORCID,Jellesen Morten Stendahl2,Rölfing Jan Duedal3ORCID,Rawat Jaideep1

Affiliation:

1. Flinders Medical Centre, Bedford Park, Australia

2. Department of Civil and Mechanical Engineering, Technical University of Denmark, Lyngby, Denmark

3. Children's Orthopaedics and Reconstruction, Aarhus University Hospital, Aarhus, Denmark

Abstract

Case: A 15-year-old girl known with osteogenesis imperfecta presented with left femoral pain. She had been treated with multiple Fassier-Duval intramedullary nails, which were still in situ. Radiographic imaging demonstrated focal osteolysis and periosteal reaction at the telescopic junction of the rod in the distal femur. She underwent implant removal. Intraoperative sampling demonstrating acute sterile inflammation and presence of brownish colored particles consistent with metal debris and osteolysis. Explant analysis confirmed corrosion of the stainless-steel telescopic nail as the underlying cause. Conclusion: Osteolysis and periosteal reaction because of corrosion should be considered in conjunction with other more common causes of pain, such as fracture or infection, in patients treated with telescopic intramedullary nails.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Reference23 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3